Performance of electrical impedance tomography in detecting regional tidal volumes during one-lung ventilation.

S Pulletz1, G Elke, G Zick, D Schädler, J Scholz, N Weiler, I Frerichs.   

Abstract

BACKGROUND: Electrical impedance tomography (EIT) is becoming a new medical imaging modality for continuous monitoring of regional lung function in the intensive care unit or operating room. The aim of our study was to evaluate the performance of EIT in detecting regional tidal volumes in patients during volume-controlled mechanical ventilation of one or both lungs.
METHODS: Ten adult patients undergoing elective thoracic surgery were included. EIT measurements were performed with the Goe-MF II EIT system. Data were collected before surgery during ventilation of both, the right and left lungs. Tidal volumes of 800 and 400 ml were applied during bilateral and unilateral ventilation, respectively.
RESULTS: Ventilation-related impedance changes determined in the whole chest cross-section during the right and left lung ventilation did not significantly differ from each other and were equal to 47.6+/-5.6% and 48.5+/-7.8% (mean+/-SD) of the value determined during bilateral ventilation. During unilateral ventilation, EIT clearly separated the ventilated and non-ventilated lung regions; nevertheless, ventilation-related impedance changes were also detected at the non-ventilated sides in areas corresponding to 3.4+/-4.1% and 12.4+/-6.9% of the scan halves during ventilation of the left and right lung, respectively. Changes in global tidal volumes were adequately detected by EIT during both bilateral and unilateral lung ventilation.
CONCLUSION: Although good separation of the ventilated and non-ventilated sides of the chest was possible, the data indicate that reliable quantification of regional tidal volumes during asymmetric or inhomogeneous distribution patterns requires regions-of-interest analysis.

Entities:  

Mesh:

Year:  2008        PMID: 18840115     DOI: 10.1111/j.1399-6576.2008.01706.x

Source DB:  PubMed          Journal:  Acta Anaesthesiol Scand        ISSN: 0001-5172            Impact factor:   2.105


  5 in total

1.  Evaluation of an electrical impedance tomography-based Global Inhomogeneity Index for pulmonary ventilation distribution.

Authors:  Zhanqi Zhao; Knut Möller; Daniel Steinmann; Inéz Frerichs; Josef Guttmann
Journal:  Intensive Care Med       Date:  2009-08-04       Impact factor: 17.440

2.  Regional distribution of blood volume within the preterm infant thorax during synchronised mechanical ventilation.

Authors:  Hazel R Carlisle; Ruth K Armstrong; Peter G Davis; Andreas Schibler; Inéz Frerichs; David G Tingay
Journal:  Intensive Care Med       Date:  2010-09-21       Impact factor: 17.440

3.  Functional Regions of Interest in Electrical Impedance Tomography: A Secondary Analysis of Two Clinical Studies.

Authors:  Tobias Becher; Barbara Vogt; Matthias Kott; Dirk Schädler; Norbert Weiler; Inéz Frerichs
Journal:  PLoS One       Date:  2016-03-24       Impact factor: 3.240

4.  Chest electrical impedance tomography examination, data analysis, terminology, clinical use and recommendations: consensus statement of the TRanslational EIT developmeNt stuDy group.

Authors:  Inéz Frerichs; Marcelo B P Amato; Anton H van Kaam; David G Tingay; Zhanqi Zhao; Bartłomiej Grychtol; Marc Bodenstein; Hervé Gagnon; Stephan H Böhm; Eckhard Teschner; Ola Stenqvist; Tommaso Mauri; Vinicius Torsani; Luigi Camporota; Andreas Schibler; Gerhard K Wolf; Diederik Gommers; Steffen Leonhardt; Andy Adler
Journal:  Thorax       Date:  2016-09-05       Impact factor: 9.139

5.  Electrical impedance tomography during major open upper abdominal surgery: a pilot-study.

Authors:  Maximilian S Schaefer; Viktoria Wania; Bea Bastin; Ursula Schmalz; Peter Kienbaum; Martin Beiderlinden; Tanja A Treschan
Journal:  BMC Anesthesiol       Date:  2014-07-05       Impact factor: 2.217