Literature DB >> 18827589

Does a volar locking plate provide equivalent stability as a dorsal nonlocking plate in a dorsally comminuted distal radius fracture?: a biomechanical study.

Utku Kandemir1, Amir Matityahu, Rohan Desai, Christian Puttlitz.   

Abstract

OBJECTIVES: The purpose of this study was to compare the fixation afforded by a dorsal nonlocking plate with a volar locking plate in a fracture model simulating an extra-articular distal radius fracture with dorsal comminution (OTA [Orthopaedic Trauma Association] type 23-A3.2).
METHODS: In 10 matched pairs of fresh-frozen cadaveric arms, a comminuted extra-articular dorsally unstable distal radius fracture (OTA type 23-A3.2) was created. The fractures were fixed with either dorsally placed nonlocking T-plate or volarly placed locking plate within matched pairs. The precycling stiffness with axial and torsional loading of the specimens was determined. The specimens were then loaded axially for 5000 cycles, and postcycling axial and torsional stiffness and load to failure were determined.
RESULTS: The mean axial and torsional stiffness before and after cyclic loading of fractures stabilized with dorsal nonlocking plate was not significantly different than fractures fixed with volar locking plate. Although the mean load to failure was greater for the volar locking plate group than dorsal nonlocking plate group, the difference was not significant.
CONCLUSIONS: This study suggests that the fixation obtained with volar locking plates is as stable as fixation with a dorsal plate in acute healing period and can withstand the functional demands of the immediate postoperative period in dorsally comminuted unstable extra-articular distal radius fractures. Elimination of dorsal tendinopathy by using volar locking plates may lead to fewer long-term complications. Locking plates provided better stability in specimens with osteoporosis.

Entities:  

Mesh:

Year:  2008        PMID: 18827589     DOI: 10.1097/BOT.0b013e318186006f

Source DB:  PubMed          Journal:  J Orthop Trauma        ISSN: 0890-5339            Impact factor:   2.512


  17 in total

1.  The effects of screw length on stability of simulated osteoporotic distal radius fractures fixed with volar locking plates.

Authors:  Lindley B Wall; Michael D Brodt; Matthew J Silva; Martin I Boyer; Ryan P Calfee
Journal:  J Hand Surg Am       Date:  2012-02-02       Impact factor: 2.230

2.  Damage in a Distal Radius Fracture Model Treated With Locked Volar Plating After Simulated Postoperative Loading.

Authors:  Christina Salas; Justin A Brantley; James Clark; Mahmoud Reda Taha; Orrin B Myers; Deana Mercer
Journal:  J Hand Surg Am       Date:  2018-02-06       Impact factor: 2.230

3.  Multidirectional volar fixed-angle plating using cancellous locking screws for distal radius fractures--evaluation of three screw configurations in an extra-articular fracture model.

Authors:  Patrick Weninger; Enrico Dall'Ara; Herwig Drobetz; Wolfgang Nemec; Markus Figl; Heinz Redl; Harald Hertz; Philippe Zysset
Journal:  Wien Klin Wochenschr       Date:  2011-01       Impact factor: 1.704

4.  Necessity of acceptable radiologic alignment by preoperative closed reduction for unstable distal radius fractures treated with volar locking plates.

Authors:  Hyoung-Seok Jung; Kwang-Jin Chun; Jae Yoon Kim; Jae-Sung Lee
Journal:  Eur J Trauma Emerg Surg       Date:  2020-02-19       Impact factor: 3.693

5.  [Unidirectional versus multidirectional palmar locking osteosynthesis of unstable distal radius fractures: comparative analysis with LDR 2.4 mm versus 2.7 mm matrix-Smartlock].

Authors:  M Hakimi; P Jungbluth; S Gehrmann; J Nowak; J Windolf; M Wild
Journal:  Unfallchirurg       Date:  2010-03       Impact factor: 1.000

6.  [Biomechanics of distal radius fractures : Basics principles and GPS treatment strategy for locking plate osteosynthesis].

Authors:  M Gabl; R Arora; G Schmidle
Journal:  Unfallchirurg       Date:  2016-09       Impact factor: 1.000

7.  The role of locking technology in the upper extremity.

Authors:  Siddharth B Joglekar; Asif M Ilyas
Journal:  J Hand Microsurg       Date:  2010-01-08

8.  The Unstable Distal Radius Fracture-How Do We Define It? A Systematic Review.

Authors:  Monique M J Walenkamp; Lara M Vos; Simon D Strackee; J Carel Goslings; Niels W L Schep
Journal:  J Wrist Surg       Date:  2015-11

9.  The influence of distal screw length on the primary stability of volar plate osteosynthesis--a biomechanical study.

Authors:  Sebastian F Baumbach; Alexander Synek; Hannes Traxler; Wolf Mutschler; Dieter Pahr; Yan Chevalier
Journal:  J Orthop Surg Res       Date:  2015-09-08       Impact factor: 2.359

10.  Assessment of a novel biomechanical fracture model for distal radius fractures.

Authors:  Sebastian Felix Baumbach; Enrico Dall'Ara; Patrick Weninger; Anna Antoni; Hannes Traxler; Martin Dörr; Philippe K Zysset
Journal:  BMC Musculoskelet Disord       Date:  2012-12-18       Impact factor: 2.362

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.