Elevated serum concentrations of cobalt, chromium, nickel, and manganese after metal-on-metal alloarthroplasty of the hip: a prospective randomized study.

Henrik Dahlstrand1, André Stark, Lucas Anissian, Nils P Hailer.   

Abstract

In this prospective randomized study, we investigate metal ion concentrations and clinical outcome 2 years after metal-on-metal (28 patients) or metal-on-polyethylene (26 patients) hip arthroplasty with 28-mm modular heads. Metal ion concentrations in patient serum were analyzed by high-resolution plasma mass spectrometry. The clinical outcome was almost identical in both groups with respect to the Harris hip score and the Medical Outcome Study Short Form-36; and on plain radiography, no signs of loosening occurred in any group. In the metal-on-metal group, concentrations of all investigated ions increased significantly when compared with preoperative values: cobalt, 15.3-fold (95% confidence interval [CI], 9.4-21.2); chromium, 5.2-fold (CI, 3.5-7.0); nickel, 2.1-fold (CI, 1.2-3.0); and manganese, 1.6-fold (CI, 1.3-2.0). In the metal-on-polyethylene group, a smaller but significant increase in the concentrations of cobalt and nickel occurred, whereas concentrations of the other ions did not change significantly.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18823744     DOI: 10.1016/j.arth.2008.07.019

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  15 in total

1.  Low incidence of groin pain and early failure with large metal articulation total hip arthroplasty.

Authors:  John B Meding; Lindsey K Meding; E Michael Keating; Michael E Berend
Journal:  Clin Orthop Relat Res       Date:  2012-02       Impact factor: 4.176

2.  Perivascular lymphocytic infiltration is not limited to metal-on-metal bearings.

Authors:  Vincent Y Ng; Adolph V Lombardi; Keith R Berend; Michael D Skeels; Joanne B Adams
Journal:  Clin Orthop Relat Res       Date:  2011-02       Impact factor: 4.176

Review 3.  Surgical management of hip osteoarthritis.

Authors:  Rajiv Gandhi; Anthony V Perruccio; Nizar N Mahomed
Journal:  CMAJ       Date:  2013-10-21       Impact factor: 8.262

Review 4.  Metal-on-Metal Bearing: Is This the End of the Line? We Do Not Think So.

Authors:  Henri Migaud; Sophie Putman; Antoine Combes; Charles Berton; Donatien Bocquet; Laurent Vasseur; Julien Girard
Journal:  HSS J       Date:  2012-09-11

Review 5.  Dutch guideline on total hip prosthesis.

Authors:  Bart A Swierstra; Anton M J S Vervest; Geert H I M Walenkamp; B Wim Schreurs; Pieter T J Spierings; Ide C Heyligers; Job L C van Susante; Harmen B Ettema; Mariette J Jansen; Pim J Hennis; Janneke de Vries; Sabrina B Muller-Ploeger; Margreet A Pols
Journal:  Acta Orthop       Date:  2011-10       Impact factor: 3.717

Review 6.  Comparative assessment of implantable hip devices with different bearing surfaces: systematic appraisal of evidence.

Authors:  Art Sedrakyan; Sharon-Lise T Normand; Stefan Dabic; Samantha Jacobs; Stephen Graves; Danica Marinac-Dabic
Journal:  BMJ       Date:  2011-11-29

7.  Elevation of circulating HLA DR(+) CD8(+) T-cells and correlation with chromium and cobalt concentrations 6 years after metal-on-metal hip arthroplasty.

Authors:  Nils P Hailer; Roman A Blaheta; Henrik Dahlstrand; André Stark
Journal:  Acta Orthop       Date:  2010-12-29       Impact factor: 3.717

Review 8.  Metal ion concentrations in body fluids after implantation of hip replacements with metal-on-metal bearing--systematic review of clinical and epidemiological studies.

Authors:  Albrecht Hartmann; Franziska Hannemann; Jörg Lützner; Andreas Seidler; Hans Drexler; Klaus-Peter Günther; Jochen Schmitt
Journal:  PLoS One       Date:  2013-08-07       Impact factor: 3.240

Review 9.  Chromium and cobalt ion concentrations in blood and serum following various types of metal-on-metal hip arthroplasties: a literature overview.

Authors:  Christopher Jantzen; Henrik L Jørgensen; Benn R Duus; Sune L Sporring; Jes B Lauritzen
Journal:  Acta Orthop       Date:  2013-04-18       Impact factor: 3.717

Review 10.  Increased Mortality in Metal-on-Metal versus Non-Metal-on-Metal Primary Total Hip Arthroplasty at 10 Years and Longer Follow-Up: A Systematic Review and Meta-Analysis.

Authors:  B G Pijls; J M T A Meessen; J W Schoones; M Fiocco; H J L van der Heide; A Sedrakyan; R G H H Nelissen
Journal:  PLoS One       Date:  2016-06-13       Impact factor: 3.240

View more