| Literature DB >> 18816871 |
Rémi Souchon1, Rares Salomir, Olivier Beuf, Laurent Milot, Denis Grenier, Denis Lyonnet, Jean-Yves Chapelon, Olivier Rouvière.
Abstract
The purpose of our study was to assess the feasibility of using ultrasound radiation force as a safe vibration source for transient MR elastography (t-MRE). We present a theoretical framework to predict the phase shift of the complex MRE signal, the temperature elevation due to ultrasound, and safety indicators (I(SPPA), I(SPTA), MI). Next, we report wave images acquired in porcine liver samples in vitro. MR thermometry was used to estimate the temperature elevation induced by ultrasound. Finally, we discuss the implications of our results with regard to the feasibility of using radiation force for t-MRE in a clinical setting, and a specific echo-planar imaging (EPI) MRE sequence is proposed. (c) 2008 Wiley-Liss, Inc.Entities:
Mesh:
Year: 2008 PMID: 18816871 DOI: 10.1002/mrm.21718
Source DB: PubMed Journal: Magn Reson Med ISSN: 0740-3194 Impact factor: 4.668