| Literature DB >> 18815080 |
Russell Handy1, Dan Trepanier, Grace Scott, Robert Foster, Derrick Freitag.
Abstract
A rapid, accurate, and reproducible liquid chromatography electrospray tandem mass spectrometry (LC/ESI-MS/MS) method was developed and validated for the therapeutic drug monitoring of voclosporin in human whole blood. Sample aliquots of 100muL were processed utilizing a protein precipitation procedure that contained a mixture of methanol, 0.2M ZnSO(4), and deuterated voclosporin internal standard. Supernatant was injected onto a Zorbax SB-C8, 2.1x12.5mm column (at 60 degrees C), and washed with water-acetonitrile, supplemented with 0.02% glacial acetic acid and 0.02mM sodium acetate, to remove poorly retained components. After washing, water-MeOH (with 0.02% glacial acetic acid and 0.02mM sodium acetate) was used to elute the voclosporin and internal standard to the Applied Biosystems/MDS-Sciex API3000 mass spectrometer for detection in multiple reaction monitoring. Analytical performance was assessed in the range of 1-200ng/ml in whole blood. This method has been used to quantify concentrations of voclosporin in whole blood from healthy volunteers participating in a pharmacokinetic study.Entities:
Mesh:
Substances:
Year: 2008 PMID: 18815080 DOI: 10.1016/j.jchromb.2008.08.023
Source DB: PubMed Journal: J Chromatogr B Analyt Technol Biomed Life Sci ISSN: 1570-0232 Impact factor: 3.205