Literature DB >> 18814131

Variations in breast doses for an automatic mammography unit.

Doğan Bor1, Selma Tükel, Turan Olgar, Elif Aydin.   

Abstract

PURPOSE: To assess variations of glandular doses for a group of patients when different dose modes are selected for a specific system.
MATERIALS AND METHODS: All measurements were obtained with a Senographe DMR mammography unit (GE Medical Systems). Automatic exposure control with either contrast or standard mode was routinely used in patient examinations. Entrance surface air kerma (ESAK) values were estimated from the postexposure mAs and from the recorded data. Subsequently the mean glandular dose (MGD) for each view was calculated using the conversion factors assuming 50% glandular and 50% adipose tissue composition.
RESULTS: The average MGD for the right craniocaudal view for all beam qualities was 1.65 mGy, and 46.7 mm was the average compressed breast thickness for this view. Average MGDs were 1.61, 1.76, and 1.35 mGy for Mo-Mo, Mo-Rh, and Rh-Rh anode filter selections, respectively. Conversely, 2.18 and 1.47 of breast doses were measured for contrast and standard dose modes at the most often used (Mo-Mo) anode filter selection.
CONCLUSION: Breast dose measurement techniques with the standard phantom or on real patients are easy to implement for the mammography users. Obtaining ESAK versus breast thickness relationship is useful for retrospective dose evaluation; better correlation can be established (67% vs. 52%) if they are generated separately for each dose modes.

Entities:  

Mesh:

Year:  2008        PMID: 18814131

Source DB:  PubMed          Journal:  Diagn Interv Radiol        ISSN: 1305-3825            Impact factor:   2.630


  4 in total

Review 1.  Reconstruction of absorbed doses to fibroglandular tissue of the breast of women undergoing mammography (1960 to the present).

Authors:  Isabelle Thierry-Chef; Steven L Simon; Robert M Weinstock; Deukwoo Kwon; Martha S Linet
Journal:  Radiat Res       Date:  2011-10-12       Impact factor: 2.841

2.  Radiation dose affected by mammographic composition and breast size: first application of a radiation dose management system for full-field digital mammography in Korean women.

Authors:  Ji Eun Baek; Bong Joo Kang; Sung Hun Kim; Hyun Sil Lee
Journal:  World J Surg Oncol       Date:  2017-02-02       Impact factor: 2.754

3.  Hybrid Detection of Breast Abnormalities Based on Contrast Agents: Introducing a Proof of Concept from a Physics Perspective.

Authors:  Panagiotis Liaparinos
Journal:  Sensors (Basel)       Date:  2022-10-03       Impact factor: 3.847

4.  Dose reduction in automatic optimization parameter of full field digital mammography: breast phantom study.

Authors:  Myung-Su Ko; Hak Hee Kim; Joo Hee Cha; Hee Jung Shin; Jeoung Hyun Kim; Min Jeong Kim
Journal:  J Breast Cancer       Date:  2013-03-31       Impact factor: 3.588

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.