Multiecho, spin-echo sequence to eliminate unwanted echoes.

Y Zur1, X Zou, L J Neuringer.   

Abstract

Two types of artifacts--a mirror-reversed image about the phase-encode direction and a wave-like intensity variation across the image--may appear when using a conventional multiecho spin-echo sequence. We describe and demonstrate a new method, which eliminates such artifacts, using only one excitation per phase-encoding step and without the need for spoiler gradients.

Mesh:

Year:  1991        PMID: 1881335     DOI: 10.1002/mrm.1910190239

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668