Gaseous microemboli detection in a simulated pediatric CPB circuit using a novel ultrasound system.

Akemi Miller1, Shigang Wang, John L Myers, Akif Undar.   

Abstract

The objective of this study was to evaluate the effect of flow rate and perfusion mode on the delivery of gaseous microemboli in a simulated pediatric cardiopulmonary bypass (CPB) circuit with an open arterial filter purge line using a novel ultrasound detection system. The circuit was primed with 450 ml fresh, heparinized bovine blood plus 200 ml Lactated Ringer's solution (total volume 650 ml, corrected Hct 25%). After the injection of 5 ml air into the venous line, an Emboli Detection and Classification (EDAC) Quantifier (Luna Innovations, Inc., Roanoke, VA) was used to simultaneously record microemboli counts at postpump, postoxygenator, and postarterial filter sites. Trials were conducted at four different flow rates (500, 750, 1,000, 1,250 ml/min) and two perfusion modes (pulsatile, nonpulsatile). Microemboli counts uniformly increased with increasing pump flow rates. In all trials, the majority of gaseous microemboli detected in the simulated pediatric CPB circuit were <20 microm in diameter. At the lowest flow rate tested (500 ml/min), all microemboli (>10 microm) were cleared from the circuit by the oxygenator and arterial filter. Clearance efficiency was decreased at higher flow rates (750-1,250 ml/min). Over 98% of microemboli detected at the postoxygenator site were <40 microm in diameter. In general, pulsatile flow delivered more microemboli to the circuit at postpump and postoxygenator sites than nonpulsatile flow. The results of this study confirmed that entrained air from the venous line could be delivered to the systemic circulation (as represented by our pediatric pseudo patient) at flow rates from 750 to 1,250 ml/min, despite the presence of an arterial filter and open arterial filter purge line. All of the microemboli distal to the arterial filter were smaller than the conventional detectable level of 40 microm.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18812742     DOI: 10.1097/MAT.0b013e318186d32a

Source DB:  PubMed          Journal:  ASAIO J        ISSN: 1058-2916            Impact factor:   2.872


  2 in total

1.  Use of the Reamer/Irrigator/Aspirator Decreases Carotid and Cranial Embolic Events in a Canine Model.

Authors:  Anna N Miller; Dwight Deal; James Green; Timothy Houle; William Brown; Clara Thore; David Stump; Lawrence X Webb
Journal:  J Bone Joint Surg Am       Date:  2016-04-20       Impact factor: 5.284

2.  In-Vitro Evaluation of Two Types of Neonatal Oxygenators in Handling Gaseous Microemboli and Maintaining Optimal Hemodynamic Stability During Cardiopulmonary Bypass.

Authors:  Neelima Marupudi; Shigang Wang; Luiz Fernando Canêo; Fabio Biscegli Jatene; Allen R Kunselman; Akif Undar
Journal:  Braz J Cardiovasc Surg       Date:  2016 Sep-Oct
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.