Ventilation patterns influence airway secretion movement.

Marcia S Volpe1, Alexander B Adams, Marcelo B P Amato, John J Marini.   

Abstract

BACKGROUND: Retention of airway secretions is a common and serious problem in ventilated patients. Treating or avoiding secretion retention with mucus thinning, patient-positioning, airway suctioning, or chest or airway vibration or percussion may provide short-term benefit.
METHODS: In a series of laboratory experiments with a test-lung system we examined the role of ventilator settings and lung-impedance on secretion retention and expulsion. Known quantities of a synthetic dye-stained mucus simulant with clinically relevant properties were injected into a transparent tube the diameter of an adult trachea and exposed to various mechanical-ventilation conditions. Mucus-simulant movement was measured with a photodensitometric technique and examined with image-analysis software. We tested 2 mucus-simulant viscosities and various peak flows, inspiratory/expiratory flow ratios, intrinsic positive end-expiratory pressures, ventilation waveforms, and impedance values.
RESULTS: Ventilator settings that produced flow bias had a major effect on mucus movement. Expiratory flow bias associated with intrinsic positive end-expiratory pressure generated by elevated minute ventilation moved mucus toward the airway opening, whereas intrinsic positive end-expiratory pressure generated by increased airway resistance moved the mucus toward the lungs. Inter-lung transfer of mucus simulant occurred rapidly across the "carinal divider" between interconnected test lungs set to radically different compliances; the mucus moved out of the low-compliance lung and into the high-compliance lung.
CONCLUSIONS: The movement of mucus simulant was influenced by the ventilation pattern and lung impedance. Flow bias obtained with ventilator settings may clear or embed mucus during mechanical ventilation.

Entities:  

Mesh:

Year:  2008        PMID: 18811989

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  17 in total

1.  Efficacy of Respiratory Physiotherapy Interventions for Intubated and Mechanically Ventilated Adults with Pneumonia: A Systematic Review and Meta-Analysis.

Authors:  Lisa van der Lee; Anne-Marie Hill; Angela Jacques; Shane Patman
Journal:  Physiother Can       Date:  2021       Impact factor: 1.037

2.  Effects of chest wall compression on expiratory flow rates in patients with chronic obstructive pulmonary disease.

Authors:  Masafumi Nozoe; Kyoshi Mase; Tomoyuki Ogino; Shigefumi Murakami; Sachie Takashima; Kazuhisa Domen
Journal:  Braz J Phys Ther       Date:  2016-03-15       Impact factor: 3.377

3.  Effects of manual hyperinflation, clinical practice versus expert recommendation, on displacement of mucus simulant: A laboratory study.

Authors:  Marcia S Volpe; Juliane M Naves; Gabriel G Ribeiro; Gualberto Ruas; Mauro R Tucci
Journal:  PLoS One       Date:  2018-02-12       Impact factor: 3.240

Review 4.  Criteria for enhancing mucus transport: a systematic scoping review.

Authors:  Alison Pieterse; Susan D Hanekom
Journal:  Multidiscip Respir Med       Date:  2018-07-06

5.  Reply to: Comparison of bronchial hygiene techniques in mechanically ventilated patients: a randomized clinical trial.

Authors:  Wagner da Silva Naue; Bruno Barcelos Herve; Fernando Nataniel Vieira; Gracieli Nadalon Deponti; Luciane de Fraga Martins; Alexandre Simões Dias; Silvia Regina Rios Vieira
Journal:  Rev Bras Ter Intensiva       Date:  2019 Oct-Dec

6.  Acute effects of ventilator hyperinflation with increased inspiratory time on respiratory mechanics: randomized crossover clinical trial.

Authors:  Luciano Matos Chicayban
Journal:  Rev Bras Ter Intensiva       Date:  2019-10-14

7.  Effects of manual chest compression on expiratory flow bias during the positive end-expiratory pressure-zero end-expiratory pressure maneuver in patients on mechanical ventilation.

Authors:  Ana Carolina Otoni Oliveira; Daiane Menezes Lorena; Lívia Corrêa Gomes; Bianca Lorrane Reges Amaral; Márcia Souza Volpe
Journal:  J Bras Pneumol       Date:  2019-03-11       Impact factor: 2.624

Review 8.  Benefits and risks of manual hyperinflation in intubated and mechanically ventilated intensive care unit patients: a systematic review.

Authors:  Frederique Paulus; Jan M Binnekade; Margreeth B Vroom; Marcus J Schultz
Journal:  Crit Care       Date:  2012-08-03       Impact factor: 9.097

9.  Outcome measures for manual lung hyperinflation: not there yet!

Authors:  George Ntoumenopoulos
Journal:  Crit Care       Date:  2012-11-08       Impact factor: 9.097

10.  Experimental study on the efficiency and safety of the manual hyperinflation maneuver as a secretion clearance technique.

Authors:  Tatiana de Arruda Ortiz; Germano Forti; Márcia Souza Volpe; Carlos Roberto Ribeiro Carvalho; Marcelo Brito Passos Amato; Mauro Roberto Tucci
Journal:  J Bras Pneumol       Date:  2013 Mar-Apr       Impact factor: 2.624

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.