Abstract

The optimum performance of an optical oxygen sensor based on polysulfone (PSF)/[Ru(II)-Tris(4,7-diphenyl-1,10-phenanthroline)] octylsulfonate (Ru(dpp)OS) was checked by carefully tuning the parameters affecting the membrane preparation. In particular, membranes having thickness ranging between 0.2 and 8.0 microm with various luminophore concentrations were prepared by dip-coating and tested. The membrane thickness was controlled by tuning the solution viscosity, and was measured both by secondary ion mass spectrometry (SIMS) and by visible spectroscopy (Vis). Luminescence-quenching-based calibration was a single value of the Stern-Volmer constant (K'SV) for membranes containing up to 20 mmol Ru(dpp) g-1 PSF (1.35 microm average thickness). The K'SV value decreased for larger concentration. The highest sensitivity was obtained with membrane thickness around 1.6 microm, having a response time close to 1 s. Thicker membranes exhibited an emission saturation effect and were characterized by longer response time. The K'SV behavior was interpreted on the basis of a mathematical approach accounting for the contribution of luminescence lifetime (tau0), oxygen diffusion coefficient (DO2) and oxygen solubility inside the membrane (sO2) establishing the role of all of them and allowing their experimental determination. Moreover, a simple experimental way to estimate K'SV without needing calibration was proposed. It was based either on the light emission asymmetry or on the percent variation of light emission on passing from pure nitrogen to pure oxygen.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18809079     DOI: 10.1016/j.aca.2008.08.019

Source DB:  PubMed          Journal:  Anal Chim Acta        ISSN: 0003-2670            Impact factor:   6.558


  3 in total

1.  Indicators for optical oxygen sensors.

Authors:  Michela Quaranta; Sergey M Borisov; Ingo Klimant
Journal:  Bioanal Rev       Date:  2012-11-24

2.  A Flexible Optical pH Sensor Based on Polysulfone Membranes Coated with pH-Responsive Polyaniline Nanofibers.

Authors:  Nedal Abu-Thabit; Yunusa Umar; Elaref Ratemi; Ayman Ahmad; Faraj Ahmad Abuilaiwi
Journal:  Sensors (Basel)       Date:  2016-06-27       Impact factor: 3.576

3.  Study and Development of a Fluorescence Based Sensor System for Monitoring Oxygen in Wine Production: The WOW Project.

Authors:  Nicola Trivellin; Diego Barbisan; Denis Badocco; Paolo Pastore; Gaudenzio Meneghesso; Matteo Meneghini; Enrico Zanoni; Giuseppe Belgioioso; Angelo Cenedese
Journal:  Sensors (Basel)       Date:  2018-04-07       Impact factor: 3.576

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.