| Literature DB >> 18805072 |
Claudia Clavijo1, Jamie Bendrick-Peart, Yan Ling Zhang, Gillian Johnson, Antje Gasparic, Uwe Christians.
Abstract
To support animal studies and clinical pharmacokinetic trials, we developed and validated an automated, specific and highly sensitive LC-MS/MS method for the quantification of naltrexone and 6beta-naltrexol in the same run. In human plasma, the assay had a lower limit of quantitation of only 5pg/mL. This was of critical importance to follow naltrexone pharmacokinetics during its terminal elimination phase. The assay had the following key performance characteristics for naltrexone in human plasma: range of reliable quantification: 0.005-100ng/mL (r2>0.99), inter-day accuracy (0.03ng/mL): 103.7% and inter-day precision: 10.1%. There were no ion suppression, matrix interferences or carry-over.Entities:
Mesh:
Substances:
Year: 2008 PMID: 18805072 DOI: 10.1016/j.jchromb.2008.08.021
Source DB: PubMed Journal: J Chromatogr B Analyt Technol Biomed Life Sci ISSN: 1570-0232 Impact factor: 3.205