Evaluation of the accuracy and reliability of two 3-dimensional sonography methods in volume measurement of small structures: an in vitro phantom study.

Brian C W Kot1, Damian M H Sin, Michael Ying.   

Abstract

PURPOSE: To evaluate the accuracy and reliability of two 3-dimensional (3D) ultrasound imaging and measurement techniques in volume measurements using custom-made phantoms.
METHODS: A total of 20 phantoms with irregular-shaped test objects of known volume (2-9.7 ml) were constructed. The phantoms were scanned using an automated mechanical 3D sonography technique and a free-hand 3D sonography technique. The volumes of the test objects were measured with "parallel planes" and "rotating planes" techniques, respectively. The measured volumes were compared with the actual volumes of the test objects. To evaluate inter- and intraoperator measurement variability, the phantoms were scanned twice by 2 different operators.
RESULTS: Both the automated mechanical and the free-hand 3D sonography techniques were accurate and reliable. Automated mechanical 3D sonography with the parallel planes technique (accuracy, 81.5-83.4%; reproducibility, 91.1%; repeatability, 98.8-99.1%) was slightly more accurate and reliable than the free-hand rotating planes technique (accuracy, 74.7-84.2%; reproducibility, 88.4%; repeatability, 97.3-98%), but the differences were not statistically significant.
CONCLUSION: Both the automated and freehand volume measurement techniques evaluated in this study are accurate and reliable. (c) 2008 Wiley Periodicals, Inc.

Mesh:

Year:  2009        PMID: 18803314     DOI: 10.1002/jcu.20525

Source DB:  PubMed          Journal:  J Clin Ultrasound        ISSN: 0091-2751            Impact factor:   0.910


  3 in total

1.  The prevalence and significance of renal perfusion defects in early kidney transplants quantified using 3D contrast enhanced ultrasound (CEUS).

Authors:  Ben Stenberg; M Wilkinson; S Elliott; N Caplan
Journal:  Eur Radiol       Date:  2017-06-07       Impact factor: 5.315

2.  Superb microvascular imaging (SMI) findings of splenic artery pseudoaneurysm: a report of two cases.

Authors:  Yumiko Yamanaka; Hideaki Ishida; Hiroko Naganuma; Tomoya Komatsuda; Hideaki Miyazawa; Takaharu Miyauchi; Satoshi Takahashi; Tomoki Tozawa; Katsuhiko Enomoto
Journal:  J Med Ultrason (2001)       Date:  2018-01-30       Impact factor: 1.314

3.  Assessing the accuracy and reliability of ultrasonographic three-dimensional parathyroid volume measurement in a patient with secondary hyperparathyroidism: a comparison with the two-dimensional conventional method.

Authors:  Sung-Hye You; Sang-Il Suh; In-Seon Ryoo; Gyu-Ri Son; Young-Hen Lee; Hyung Suk Seo; Nam Joon Lee; Hae Young Seol
Journal:  Ultrasonography       Date:  2016-06-08
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.