Literature DB >> 1879856

Control of specific absorption rate distribution using capacitive electrodes and inductive aperture-type applicators: implications for radiofrequency hyperthermia.

H Kato1, J W Hand, M V Prior, M Furukawa, O Yamamoto, T Ishida.   

Abstract

A method of controlling the specific absorption rate (SAR) distribution in radiofrequency hyperthermia is proposed. The superposition of the current density associated with capacitively coupled electrodes and that associated with H-field coupled inductive aperture-type applicators modifies the actual current distribution in the heating material. Using a two-dimensional finite element method, we have shown that "focusing" is possible such that the SAR at the center of a phantom can be adjusted to be approximately twice that in superficial regions, even though the wavelength is considerably greater than the dimensions of the phantom.

Entities:  

Mesh:

Year:  1991        PMID: 1879856     DOI: 10.1109/10.83563

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  1 in total

Review 1.  Present and future status of noninvasive selective deep heating using RF in hyperthermia.

Authors:  H Kato; T Ishida
Journal:  Med Biol Eng Comput       Date:  1993-07       Impact factor: 2.602

  1 in total