| Literature DB >> 1879856 |
H Kato1, J W Hand, M V Prior, M Furukawa, O Yamamoto, T Ishida.
Abstract
A method of controlling the specific absorption rate (SAR) distribution in radiofrequency hyperthermia is proposed. The superposition of the current density associated with capacitively coupled electrodes and that associated with H-field coupled inductive aperture-type applicators modifies the actual current distribution in the heating material. Using a two-dimensional finite element method, we have shown that "focusing" is possible such that the SAR at the center of a phantom can be adjusted to be approximately twice that in superficial regions, even though the wavelength is considerably greater than the dimensions of the phantom.Entities:
Mesh:
Year: 1991 PMID: 1879856 DOI: 10.1109/10.83563
Source DB: PubMed Journal: IEEE Trans Biomed Eng ISSN: 0018-9294 Impact factor: 4.538