Literature DB >> 18789446

The stiffness characteristics of hybrid Ilizarov fixators.

Onder Baran1, Hasan Havitcioglu, Hasan Tatari, Berivan Cecen.   

Abstract

The use of hybrid Ilizarov models around femoral area is gaining clinical popularity lately. Hybrid systems show different mechanical properties. The purpose of this report is to examine the stiffness characteristics of the C-arch and half-pins on the hybrid Ilizarov fixators. Eight models that included standard Ilizarov and hybrid models were applied to six femoral sawbones. The distal part of fixation was composed of a two-ring frame applied identically to all bones. The difference of the configuration was at the proximal part, where half-pins with or without C-arches were either added to the proximal two-ring frame or replaced the proximal one- or two-ring frame. Osteotomy was performed in the femoral diaphysis and the bone was distracted 2cm. The stability of the system was tested with the axial compression testing machine. Displacements between the adjacent fracture sides were measured with the video extensometer in three dimensions. We found that proximal half-pin applications alone had less stiffness, but half-pins with C-arch had more stiffness than the model including only half-pins. Additional half-pins onto one- or two-ring frames had more longitudinal stiffness, but this system showed weak resistance against transverse displacement.

Entities:  

Mesh:

Year:  2008        PMID: 18789446     DOI: 10.1016/j.jbiomech.2008.07.030

Source DB:  PubMed          Journal:  J Biomech        ISSN: 0021-9290            Impact factor:   2.712


  7 in total

1.  Biomechanical properties of different external fixator frame configurations.

Authors:  R M Sellei; P Kobbe; T Dienstknecht; P Lichte; R Pfeifer; M Behrens; S Brianza; H-C Pape
Journal:  Eur J Trauma Emerg Surg       Date:  2014-08-02       Impact factor: 3.693

Review 2.  Outcomes following the treatment of bicondylar tibial plateau fractures with fine wire circular frame external fixation compared to open reduction and internal fixation: A systematic review.

Authors:  Tarek Boutefnouchet; Ayaz S Lakdawala; Panayiotis Makrides
Journal:  J Orthop       Date:  2015-02-24

3.  Configuration design and correction ability evaluation of a novel external fixator for foot and ankle deformity treated by U osteotomy.

Authors:  Shiping Zuo; Mingjie Dong; Jianfeng Li; Chunjing Tao; Run Ji
Journal:  Med Biol Eng Comput       Date:  2020-01-08       Impact factor: 2.602

4.  Biomechanical Analysis of the Behaviour at the Metaphyseal-Diaphyseal Junction of Complex Tibial Plateau Fractures Using Two Circular Fixator Configurations.

Authors:  Gracielle S Cardoso; Renato Amorim; Francisco M Penha; Françoá J Horn; Carlos Rm Roesler; Jefferson Lb Marques
Journal:  Strategies Trauma Limb Reconstr       Date:  2020 Sep-Dec

5.  The influence of a weight-bearing platform on the mechanical behavior of two Ilizarov ring fixators: tensioned wires vs. half-pins.

Authors:  Jan Gessmann; Mustafa Citak; Birger Jettkant; Thomas A Schildhauer; Dominik Seybold
Journal:  J Orthop Surg Res       Date:  2011-12-12       Impact factor: 2.359

6.  Mechanical rigidity of the Ortho-SUV frame compared to the Ilizarov frame in the correction of femoral deformity.

Authors:  Petr V Skomoroshko; Victor A Vilensky; Ahmed I Hammouda; Matt D A Fletcher; Leonid N Solomin
Journal:  Strategies Trauma Limb Reconstr       Date:  2015-02-26

7.  Efficacy of a compliant semicircular Ilizarov pin fixator module for treating infected nonunion of the femoral diaphysis.

Authors:  Ashraf A Khanfour; Mohamed M El-Sayed
Journal:  Strategies Trauma Limb Reconstr       Date:  2014-07-29
  7 in total