Literature DB >> 18783991

A dosimetric comparison of Xoft Axxent Electronic Brachytherapy and iridium-192 high-dose-rate brachytherapy in the treatment of endometrial cancer.

Adam Dickler1, Michael C Kirk, Alan Coon, Damian Bernard, Tom Zusag, Jacob Rotmensch, David E Wazer.   

Abstract

PURPOSE: This analysis was undertaken to dosimetrically compare iridium-192 high-dose-rate brachytherapy (IB) and Xoft Axxent Electronic Brachytherapy (XB; Xoft Inc., Sunnyvale, CA) in the treatment of endometrial cancer. METHODS AND MATERIALS: The planning CT scans from 11 patients previously treated with IB were used to construct hypothetical treatment plans using the source characteristics of the XB device. The mean V95, V100, and V150 (percent of the planning target volume that received 95%, 100%, and 150% of the prescription dose) were calculated. For both the bladder and rectum, the V35 (percent of the organ that received 35% of the prescription dose) and V50 (percent of the organ that received 50% of the prescription dose) were calculated for each patient using both methods of vaginal brachytherapy.
RESULTS: The mean %V95 was 99.7% vs. 99.6% (p = ns) and the mean %V100 was 99.0% vs. 99.1% (p = ns) for the IB and XB methods, respectively. The mean %V150 was 35.8% vs. 58.9% (p < 0.05) for the IB and XB methods, respectively. The mean bladder %V35 was 47.7% vs. 27.4% (p < 0.05) and the mean bladder %V50 was 26.5% vs. 15.9% (p < 0.05) for the IB and XB methods, respectively. The mean rectal %V35 was 48.3% vs. 28.3% (p < 0.05) and the mean rectal %V50 was 27.8% vs. 17.0% (p < 0.05) for the IB and XB methods, respectively.
CONCLUSIONS: The IB and XB methods of vaginal brachytherapy offer equivalent target volume coverage; however, the XB method allows increased sparing of the bladder and rectum.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18783991     DOI: 10.1016/j.brachy.2008.05.003

Source DB:  PubMed          Journal:  Brachytherapy        ISSN: 1538-4721            Impact factor:   2.362


  8 in total

Review 1.  Electronic brachytherapy--current status and future directions.

Authors:  D J Eaton
Journal:  Br J Radiol       Date:  2015-03-06       Impact factor: 3.039

2.  Comparison of Axxent-Xoft, (192)Ir and (60)Co high-dose-rate brachytherapy sources for image-guided brachytherapy treatment planning for cervical cancer.

Authors:  P N Mobit; S Packianathan; R He; C C Yang
Journal:  Br J Radiol       Date:  2015-05-21       Impact factor: 3.039

3.  Prospective multi-center trial utilizing electronic brachytherapy for the treatment of endometrial cancer.

Authors:  Adam Dickler; Mohamed Y Puthawala; John P Thropay; Ajay Bhatnagar; Gary Schreiber
Journal:  Radiat Oncol       Date:  2010-07-20       Impact factor: 3.481

4.  Use of electronic brachytherapy to deliver postsurgical adjuvant radiation therapy for endometrial cancer: a retrospective multicenter study.

Authors:  William C Dooley; John P Thropay; Gary J Schreiber; Mohamed Y Puthawala; Steven C Lane; James C Wurzer; Charles E Stewart; Gordon L Grado; Harish G Ahuja; Gary M Proulx
Journal:  Onco Targets Ther       Date:  2010-10-05       Impact factor: 4.147

5.  Dosimetric comparison of brachytherapy sources for high-dose-rate treatment of endometrial cancer: (192)Ir, (60)Co and an electronic brachytherapy source.

Authors:  Paul N Mobit; Alex Nguyen; Satyaseelan Packianathan; Rui He; Claus C Yang
Journal:  Br J Radiol       Date:  2016-01-08       Impact factor: 3.039

6.  Electronic brachytherapy as adjuvant therapy for early stage breast cancer: a retrospective analysis.

Authors:  William C Dooley; James C Wurzer; Mohamed Megahy; Gary Schreiber; Tapan Roy; Gary Proulx; Susan Laduzinsky; Steven Lane; James Dalzell; Kambiz Dowlatshahi; Dwelvin Simmons; John P Thropay; Harish Ahuja; Peter Beitsch; Randall W Holt; Charles A Lee
Journal:  Onco Targets Ther       Date:  2011-01-12       Impact factor: 4.147

7.  Comparison of tumor and normal tissue dose for accelerated partial breast irradiation using an electronic brachytherapy eBx source and an Iridium-192 source.

Authors:  Salahuddin Ahmad; Daniel Johnson; Jessica R Hiatt; D Timothy Still; Eli E Furhang; David Marsden; Frank Kearly; Damian A Bernard; Randall W Holt
Journal:  J Appl Clin Med Phys       Date:  2010-09-14       Impact factor: 2.102

8.  Treatment of cervical cancer with electronic brachytherapy.

Authors:  Sergio Lozares-Cordero; José Antonio Font-Gómez; Almudena Gandía-Martínez; Anabela Miranda-Burgos; Agustina Méndez-Villamón; David Villa-Gazulla; Verónica Alba-Escorihuela; Sara Jiménez-Puertas; Víctor González-Pérez
Journal:  J Appl Clin Med Phys       Date:  2019-06-11       Impact factor: 2.102

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.