Bilateral tension pneumothorax due to a defective anesthesia breathing circuit filter.

C E Smith1, J R Otworth, P Kaluszyk.   

Abstract

A 25-year-old healthy patient developed bilateral tension pneumothorax during general endotracheal anesthesia due to a defective anesthesia breathing circuit filter. Prompt diagnosis and treatment of the pneumothorax was facilitated by the use of capnography and pulse oximetry. The manufacturing process of the breathing circuit and filter that made this accident possible has since been corrected by the manufacturer. Anesthesiologists must be alert to the possibility of such accidents with any breathing system using bacterial filters.

Entities:  

Mesh:

Year:  1991        PMID: 1878237     DOI: 10.1016/0952-8180(91)90166-k

Source DB:  PubMed          Journal:  J Clin Anesth        ISSN: 0952-8180            Impact factor:   9.452


  2 in total

1.  Complete bacterial filter blockage by a plastic film.

Authors:  Jiapeng Huang; Kishin Dodwani; Paul Simon; Jing Zhou
Journal:  J Anaesthesiol Clin Pharmacol       Date:  2012-04

2.  Life threatening tension pneumothorax during cardiac surgery. A case report.

Authors:  A Jain; D Arora; R Juneja; Y Mehta; N Trehan
Journal:  Heart Lung Vessel       Date:  2014
  2 in total