Literature DB >> 18774548

Comparison of Fourier-domain and time-domain optical coherence tomography in the detection of band atrophy of the optic nerve.

Luciana V F Costa-Cunha1, Leonardo P Cunha, Roberto F S Malta, Mário L R Monteiro.   

Abstract

PURPOSE: To compare the ability of Fourier-domain (FD) optical coherence tomography (3D OCT-1000; Topcon, Tokyo, Japan) and time-domain (TD) OCT (Stratus; Carl Zeiss Meditec Inc, Dublin, California, USA) to detect axonal loss in eyes with band atrophy (BA) of the optic nerve.
DESIGN: Cross-sectional study.
METHODS: Thirty-six eyes from 36 patients with BA and temporal visual field (VF) defect from chiasmal compression and 36 normal eyes were studied. Subjects were submitted to standard automated perimetry and macular and retinal nerve fiber layer (RNFL) measurements were taken using 3D OCT-1000 and Stratus OCT. Receiver operating characteristic (ROC) curves were calculated for each parameter. Spearman correlation coefficients were obtained to evaluate the relationship between RNFL and macular thickness parameters and severity of VF loss. Measurements from the two devices were compared.
RESULTS: Regardless of OCT device, all RNFL and macular thickness parameters were significantly lower in eyes with BA compared with normal eyes, but no statistically significant difference was found with regard to the area under the ROC curve. Structure-function relationships were also similar for the two devices. In both groups, RNFL and macular thickness measurements were generally and in some cases significantly smaller with 3D OCT-1000 than with Stratus OCT.
CONCLUSIONS: The introduction of FD technology did not lead to better discrimination ability for detecting BA of the optic nerve compared with TD technology when using the software currently provided by the manufacturer. 3D OCT-1000 FD OCT RNFL and macular measurements were generally smaller than TD Stratus OCT measurements. Investigators should be aware of this fact when comparing measurements obtained with these two devices.

Entities:  

Mesh:

Year:  2008        PMID: 18774548     DOI: 10.1016/j.ajo.2008.07.020

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  24 in total

1.  Relationship between macular ganglion cell complex parameters and visual field parameters after tumor resection in chiasmal compression.

Authors:  Shinji Ohkubo; Tomomi Higashide; Hisashi Takeda; Eiji Murotani; Yasuhiko Hayashi; Kazuhisa Sugiyama
Journal:  Jpn J Ophthalmol       Date:  2012-01       Impact factor: 2.447

2.  Better performance of RTVue than Cirrus spectral-domain optical coherence tomography in detecting band atrophy of the optic nerve.

Authors:  Makoto Nakamura; Kumiko Ishikawa-Tabuchi; Akiyasu Kanamori; Yuko Yamada; Akira Negi
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-07-01       Impact factor: 3.117

3.  Evaluation of inner retinal layers in eyes with temporal hemianopic visual loss from chiasmal compression using optical coherence tomography.

Authors:  Mário L R Monteiro; Kenzo Hokazono; Danilo B Fernandes; Luciana V F Costa-Cunha; Rafael M Sousa; Ali S Raza; Diane L Wang; Donald C Hood
Journal:  Invest Ophthalmol Vis Sci       Date:  2014-04-24       Impact factor: 4.799

4.  Macular thickness measurements with frequency domain-OCT for quantification of axonal loss in chronic papilledema from pseudotumor cerebri syndrome.

Authors:  M L R Monteiro; C L Afonso
Journal:  Eye (Lond)       Date:  2014-01-10       Impact factor: 3.775

5.  Adaptive optics imaging of healthy and abnormal regions of retinal nerve fiber bundles of patients with glaucoma.

Authors:  Monica F Chen; Toco Y P Chui; Paula Alhadeff; Richard B Rosen; Robert Ritch; Alfredo Dubra; Donald C Hood
Journal:  Invest Ophthalmol Vis Sci       Date:  2015-01-08       Impact factor: 4.799

6.  Visual Outcomes, Visual Fields, and Optical Coherence Tomography in Paediatric Craniopharyngioma.

Authors:  Soraya Mediero; Susana Noval; Luciano Bravo-Ljubetic; Inés Contreras; Fernando Carceller
Journal:  Neuroophthalmology       Date:  2015-06-17

Review 7.  Optical coherence tomography as a marker of vision in children with optic pathway gliomas.

Authors:  Ana Banc; Cristina Stan; Ioan Stefan Florian
Journal:  Childs Nerv Syst       Date:  2017-08-26       Impact factor: 1.475

Review 8.  OCT: New perspectives in neuro-ophthalmology.

Authors:  Gema Rebolleda; Laura Diez-Alvarez; Alfonso Casado; Carmen Sánchez-Sánchez; Elisabet de Dompablo; Julio J González-López; Francisco J Muñoz-Negrete
Journal:  Saudi J Ophthalmol       Date:  2014-10-05

Review 9.  Optical coherence tomography impacts the evaluation of visual pathway tumors.

Authors:  Ana Banc; Cristina Stan; Ioan Stefan Florian
Journal:  Neurosurg Rev       Date:  2016-07-28       Impact factor: 3.042

10.  Segmentation of the optic disc in 3-D OCT scans of the optic nerve head.

Authors:  Kyungmoo Lee; Meindert Niemeijer; Mona K Garvin; Young H Kwon; Milan Sonka; Michael D Abramoff
Journal:  IEEE Trans Med Imaging       Date:  2009-09-15       Impact factor: 10.048

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.