Calibration of reference KAP-meters at SSDL and cross calibration of clinical KAP-meters.

Per O Hetland1, Eva G Friberg, Kirsti M Ovrebø, Hans H Bjerke.   

Abstract

INTRODUCTION: In the summer of 2007 the secondary standard dosimetry laboratory (SSDL) in Norway established a calibration service for reference air-kerma product meter (KAP-meter). The air-kerma area product, P(KA), is a dosimetric quantity that can be directly related to the patient dose and used for risk assessment associated with different x-ray examinations. The calibration of reference KAP-meters at the SSDL gives important information on parameters influencing the calibration factor for different types of KAP-meters. The use of reference KAP-meters calibrated at the SSDL is an easy and reliable way to calibrate or verify the P(KA) indicated by the x-ray equipment out in the clinics.
MATERIAL AND METHODS: Twelve KAP-meters were calibrated at the SSDL by use of the substitution method at five diagnostic radiation qualities (RQRs).
RESULTS: The calibration factors varied from 0.94 to 1.18. The energy response of the individual KAP-meters varied by a total of 20% between the different RQRs and the typical chamber transmission factors ranged from 0.78 to 0.91. DISCUSSION: It is important to use a calibrated reference KAP-meter and a harmonised calibration method in the P(KA) calibration in hospitals. The obtained uncertainty in the P(KA) readings is comparable with other calibration methods if the information in the calibration certificate is correct used, corrections are made and proper positioning of the KAP-chamber is performed. This will ensure a reliable estimate of the patient dose and a proper optimisation of conventional x-ray examinations and interventional procedures.

Entities:  

Mesh:

Year:  2009        PMID: 18759141     DOI: 10.1080/02841860802287124

Source DB:  PubMed          Journal:  Acta Oncol        ISSN: 0284-186X            Impact factor:   4.089


  1 in total

1.  CT image quality over time: comparison of image quality for six different CT scanners over a six-year period.

Authors:  Ana Maria A Roa; Hilde K Andersen; Anne Catrine T Martinsen
Journal:  J Appl Clin Med Phys       Date:  2015-03-08       Impact factor: 2.102

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.