Sherri L Kacinko1, Marta Concheiro-Guisan, Diaa M Shakleya, Marilyn A Huestis.
Abstract
A liquid chromatography-tandem mass spectrometry method for the simultaneous quantification of buprenorphine (BUP), norbuprenorphine (NBUP), buprenorphine glucuronide (BUP-Gluc), and norbuprenorphine glucuronide (NBUP-Gluc) in human urine was developed and fully validated. Extensive endogenous and exogenous interferences were evaluated and limits of quantification were identified empirically. Analytical ranges were 5-1,000 ng/mL for BUP and BUP-Gluc and 25-1,000 ng/mL for NBUP and NBUP-Gluc. Intra-assay and interassay imprecision were less than 17% and recovery was 93-116%. Analytes were stable at room temperature, at 4 degrees C, and for three freeze-thaw cycles. This accurate and precise assay has sufficient sensitivity and specificity for urine analysis of specimens collected from individuals treated with BUP for opioid dependence.Entities:
Mesh:
Substances:
Year: 2008 PMID: 18758763 PMCID: PMC2670341 DOI: 10.1007/s00216-008-2326-z
Source DB: PubMed Journal: Anal Bioanal Chem ISSN: 1618-2642 Impact factor: 4.142