Literature DB >> 18756699

Influence of contaminants in the stem-ball interface on the static fracture load of ceramic hip joint ball heads.

B Weisse1, C Affolter, A Stutz, G P Terrasi, S Köbel, W Weber.   

Abstract

The probability of in-vivo failure of ceramic hip joint implants is very low (0.05-0.004 per cent). Besides material flaws and overloading, improper handling during implantation may induce fractures of the ceramic ball head in the long term. This study focuses on the influence of contaminants located in the stem-ball interface and on the use of damaged metal tapers on the strength of ceramic ball heads. Mechanical tests on alumina ball heads according to the standard ISO 7206-10 were performed to identify their effect on the static fracture load. A decrease of up to 90 per cent with respect to the reference static fracture load was found when contaminants such as bone chips, soft tissue, or blood were present. Reductions of 57 per cent and 27 per cent were observed for deformed stem cross-sections (from circular to elliptical) and for flattened stems respectively, making deformed stems another influential parameter. Since any alteration of the interface between the metal taper and the ceramic ball head yields a nonuniform load introduction and hence results in stress concentrations, its presence has to be avoided.

Entities:  

Mesh:

Year:  2008        PMID: 18756699     DOI: 10.1243/09544119JEIM374

Source DB:  PubMed          Journal:  Proc Inst Mech Eng H        ISSN: 0954-4119            Impact factor:   1.617


  6 in total

1.  Primary ceramic-on-ceramic total hip arthroplasty using a 32-mm ceramic head with a titanium-alloy sleeve.

Authors:  Seung-Jae Lim; Seung-Pil Jang; Dong-Wook Kim; Young-Wan Moon; Youn-Soo Park
Journal:  Clin Orthop Relat Res       Date:  2015-12       Impact factor: 4.176

Review 2.  [The problem of fractures of ceramic heads. What should be done?].

Authors:  H R Dürr
Journal:  Orthopade       Date:  2009-08       Impact factor: 1.087

Review 3.  What the Surgeon Can Do to Reduce the Risk of Trunnionosis in Hip Arthroplasty: Recommendations from the Literature.

Authors:  Claude B Rieker; Peter Wahl
Journal:  Materials (Basel)       Date:  2020-04-21       Impact factor: 3.623

4.  Influence of Different Damage Patterns of the Stem Taper on Fixation and Fracture Strength of Ceramic Ball Heads for Total Hip Replacement.

Authors:  Danny Vogel; Jessica Hembus; Mario Jackszis; Vera Bolte; Rainer Bader
Journal:  Biomed Res Int       Date:  2020-05-13       Impact factor: 3.411

5.  Quantification of the Contact Area at the Head-Stem Taper Interface of Modular Hip Prostheses.

Authors:  Florian Witt; Julian Gührs; Michael M Morlock; Nicholas E Bishop
Journal:  PLoS One       Date:  2015-08-17       Impact factor: 3.240

6.  Are asymmetric metal markings on the cone surface of ceramic femoral heads an indication of entrapped debris?

Authors:  Sebastian Valet; Bernhard Weisse; Jakob Kuebler; Martin Zimmermann; Christian Affolter; Giovanni Pietro Terrasi
Journal:  Biomed Eng Online       Date:  2014-04-04       Impact factor: 2.819