Literature DB >> 18728382

[Evaluation of performance and false positivity of Mediace RPR test that uses a chemistry autoanalyzer].

Jaekwang Noh1, Hak Hyun Ko, Yeomin Yun, Young Sook Choi, Sang Gon Lee, Sue Shin, Kyou Sup Han, Eun Young Song.   

Abstract

BACKGROUND: We evaluated the performance and false positive rate of Mediace RPR test (Sekisui, Japan), a newly introduced nontreponemal test using a chemistry autoanalyzer.
METHODS: The sensitivity of Mediace RPR test was analyzed using sera from 50 patients with syphilis in different stages (8 primary, 7 secondary, and 35 latent), 14 sera positive with fluorescent treponemal antibody absorption (FTA-ABS) IgM, and 74 sera positive with conventional rapid plasma regain (RPR) card test (Asan, Korea) and also positive with Treponema pallidum hemagglutination (TPHA) test or FTA-ABS IgG test. The specificity was analyzed on 108 healthy blood donors. We also performed RPR card test on 302 sera that had been tested positive with Mediace RPR test and also performed TPHA or FTA-ABS IgG test to analyze the false positive rate of Mediace RPR test. A cutoff value of 0.5 R.U. (RPR unit) was used for Mediace RPR test.
RESULTS: Mediace RPR test on syphilitic sera of different stages (primary, secondary, and latent stages) and FTA-ABS IgM positive sera showed a sensitivity of 100%, 100%, 82.9% and 100%, respectively. Among the 74 sera positive with conventional RPR card test and TPHA or FTA-ABS IgG test, 55 were positive with Mediace test. The specificity of Mediace RPR test on blood donors was 97.2%. Among the 302 sera positive with Mediace RPR test, 137 sera (45.4%) were negative by RPR card and TPHA/FTA-ABS IgG tests.
CONCLUSIONS: Although the sensitivities of Mediace RPR were good for primary and secondary syphilis, due to its high negative rate of Mediace RPR over the conventional RPR positive samples, further studies are necessary whether it can replace conventional nontreponemal test for screening purpose. Moreover, in view of the high false positive rate, positive results by Mediace RPR test should be confirmed with treponemal tests.

Entities:  

Mesh:

Year:  2008        PMID: 18728382     DOI: 10.3343/kjlm.2008.28.4.312

Source DB:  PubMed          Journal:  Korean J Lab Med        ISSN: 1598-6535


  4 in total

1.  Evaluation of a new chemiluminescence immunoassay for diagnosis of syphilis.

Authors:  Xiaohui Mo; Yuelan Jin; Yang Yang; Weizhong Hu; Weiming Gu
Journal:  Eur J Med Res       Date:  2010-02-26       Impact factor: 2.175

2.  Comparison of Manual and Fully Automated AIX1000 Rapid Plasma Reagin Assays for Laboratory Diagnosis of Syphilis.

Authors:  Alan M Sanfilippo; Kristie Freeman; John L Schmitz
Journal:  J Clin Microbiol       Date:  2018-07-26       Impact factor: 5.948

3.  False-positive and false-negative COVID-19 cases: respiratory prevention and management strategies, vaccination, and further perspectives.

Authors:  Dimitra S Mouliou; Konstantinos I Gourgoulianis
Journal:  Expert Rev Respir Med       Date:  2021-04-25       Impact factor: 3.772

4.  Comparison of automated treponemal and nontreponemal test algorithms as first-line syphilis screening assays.

Authors:  Hee Jin Huh; Jae Woo Chung; Seong Yeon Park; Seok Lae Chae
Journal:  Ann Lab Med       Date:  2016-01       Impact factor: 3.464

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.