Effects of laser repetition rate on corneal tissue ablation for 193-nm excimer laser light.

Leia M Shanyfelt1, Pamela L Dickrell, Henry F Edelhauser, David W Hahn.   

Abstract

BACKGROUND AND
OBJECTIVE: The goal of the present work is to assess whether bovine corneal ablations generated at laser repetition rates of up to 400 Hz are comparable to ablations performed at rates consistent with current clinical laser systems. STUDY DESIGN/
MATERIALS AND METHODS: A combination of experiments was used to assess a comprehensive range of ablation parameters, including ablation plume dynamics via imaging and transmission, corneal ablation profiles via scanning interferometry, and high-resolution electron microscopy of collagen structure following ablation.
RESULTS: Using white-light interferometry analysis, no statistical difference was found between corneal ablation profiles created at 60 and 400 Hz, with an average rate of 0.94 microm/pulse at 60 Hz versus 0.92 microm/pulse at 400 Hz. In addition, based on plume imaging and transmission studies, the bulk ablation plume was found to dissipate on a time-scale less than the pulse-to-pulse separation for a laser repetition rate up to about 400 Hz. A persistent, diffuse gas-phase component of the ablation products was observed and concluded to be comparable at both repetition rates. Finally, SEM and TEM analysis revealed no signs of differential thermal tissue damage, including collagen fibril analysis, for laser repetition rates up to 400 Hz.
CONCLUSIONS: In summary, investigation of the relative effects of excimer laser repetition rate on the overall corneal ablation metrics revealed no measurable difference under conditions typical of clinical refractive procedures. This study suggests that increases in ArF laser repetition rates for clinical applications (up to approximately 400 Hz) appear feasible.

Entities:  

Mesh:

Year:  2008        PMID: 18727026     DOI: 10.1002/lsm.20656

Source DB:  PubMed          Journal:  Lasers Surg Med        ISSN: 0196-8092            Impact factor:   4.025


  2 in total

1.  Comparison of visual and refractive results after transepithelial and mechanical photorefractive keratectomy in myopia.

Authors:  Yusuf Yildirim; Onur Olcucu; Nese Alagoz; Alper Agca; Yalcin Karakucuk; Ahmet Demirok
Journal:  Int Ophthalmol       Date:  2017-03-27       Impact factor: 2.031

2.  Centration axis in refractive surgery.

Authors:  Samuel Arba Mosquera; Shwetabh Verma; Colm McAlinden
Journal:  Eye Vis (Lond)       Date:  2015-02-24
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.