Pulse oximetry monitoring outside the intensive care unit: progress or problem?

D L Bowton1, P E Scuderi, L Harris, E F Haponik.   

Abstract

OBJECTIVE: To evaluate the use of continuous pulse oximetry monitoring in general care units.
DESIGN: Hemoglobin oxygen saturation data collected prospectively by use of pulse oximetry with concurrent review of the medical record.
SETTING: General medical-surgical nursing units in a large, tertiary care university hospital. PATIENTS: Forty patients on two nursing units monitored with continuous, bedside pulse oximetry at the request of their primary physicians. MEASUREMENTS: All patients had continuous pulse oximetry monitoring. A research associate visited the bedside two or three times daily and recorded saturation compared with time data from the previous 8.75 hours. Patients were studied for 36 hours or until pulse oximetry monitoring was discontinued. Episodes of desaturation were counted. Patient charts were reviewed for documentation of desaturation in either nursing or physician notes. Orders adjusting oxygen therapy or other respiratory therapy within 12 hours of any desaturation episode were also recorded. MAIN
RESULTS: Thirty of the 40 patients (75%) had at least one episode of desaturation to less than 90%; 23 (58%) had at least one episode to less than 85%. Desaturation episodes were documented in nursing notes for only 33% of those patients who desaturated to less than 90% and in physician notes in only 7% of cases. Changes in respiratory therapy were ordered in 20% of patients who desaturated to less than 90% and in only 26% who desaturated to less than 85%.
CONCLUSIONS: Despite their repeated occurrence, episodes of hypoxemia were rarely documented in either nursing or physician notes. Further, even in patients who had episodic desaturation, pulse oximetry monitoring had little effect on changes in physician-directed respiratory care.

Entities:  

Mesh:

Year:  1991        PMID: 1872492     DOI: 10.7326/0003-4819-115-6-450

Source DB:  PubMed          Journal:  Ann Intern Med        ISSN: 0003-4819            Impact factor:   25.391


  3 in total

1.  Early response to pulse oximetry alarms with telemetry.

Authors:  M A Klaas; E Y Cheng
Journal:  J Clin Monit       Date:  1994-05

2.  Cost effective computerized decision support: tracking caregiver acceptance at the point of care.

Authors:  C J Wallace; S Metcalf; X Zhang; A T Kinder; L Greenway; A H Morris
Journal:  Proc Annu Symp Comput Appl Med Care       Date:  1995

3.  Pulse oximetry.

Authors: 
Journal:  Crit Care       Date:  1999       Impact factor: 9.097

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.