Positioning accuracy in stereotactic radiotherapy using a mask system with added vacuum mouth piece and stereoscopic X-ray positioning.

Jan van Santvoort1, Ruud Wiggenraad, Petra Bos.   

Abstract

PURPOSE: For cranial patients receiving stereotactic radiotherapy, we use the Exactrac stereoscopic X-ray system to optimize patient positioning. Patients are immobilized with the BrainLAB Mask System (BrainLAB, Feldkirchen, Germany). We have developed an adapter to this system that accommodates a vacuum mouth piece (VMP). Measurements with the Exactrac system have been performed to study the positioning accuracy after corrections with this system and to evaluate the accuracy of the VMP vs. the standard available upper jaw support (UJS). METHODS AND MATERIALS: Positioning results were collected for 20 patients with the UJS and 20 patients with the VMP, before treatment (1,122 fractions) and after treatment (400 fractions). For all 6 degrees of freedom the average, the random error and systematic error were calculated.
RESULTS: The average vector length before and after correction with the Exactrac system was 2.1 +/- 1.2 mm and 0.7 +/- 0.6 mm respectively for UJS and 1.7 +/- 0.7 mm and 0.4 +/- 0.4 mm for VMP. Interfraction positioning for translations was greatly improved after correction with the Exactrac system (p < 0.0005) and is better with VMP than with UJS (p = 0.005). Outliers were greatly reduced. Interfraction rotations were significantly smaller for VMP. Intrafraction errors for vertical and longitudinal translations and for rotations were smaller for the VMP.
CONCLUSIONS: Positioning correction using the Exactrac X-ray system greatly improves accuracy. Adding the VMP results in even better patient fixation and smaller rotations, making it a useful addition to the Mask System. Combined, this is a convenient and accurate alternative to invasive fixation methods.

Entities:  

Mesh:

Year:  2008        PMID: 18722276     DOI: 10.1016/j.ijrobp.2008.05.006

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  16 in total

1.  Local progression and pseudo progression after single fraction or fractionated stereotactic radiotherapy for large brain metastases. A single centre study.

Authors:  R Wiggenraad; A Verbeek-de Kanter; M Mast; R Molenaar; H B Kal; G Lycklama à Nijeholt; C Vecht; H Struikmans
Journal:  Strahlenther Onkol       Date:  2012-06-23       Impact factor: 3.621

2.  Bite-bar stabilization system for stereotactic frame application: A technical note.

Authors:  Keith Aronyk; Gregory Bowden; Angela Pickering; Kim Chandler; Samir Patel
Journal:  J Radiosurg SBRT       Date:  2020

3.  Relocatable fixation systems in intracranial stereotactic radiotherapy. Accuracy of serial CT scans and patient acceptance in a randomized design.

Authors:  A Theelen; J Martens; G Bosmans; R Houben; J J Jager; I Rutten; P Lambin; A W Minken; B G Baumert
Journal:  Strahlenther Onkol       Date:  2011-12-24       Impact factor: 3.621

4.  Semi-robotic 6 degree of freedom positioning for intracranial high precision radiotherapy; first phantom and clinical results.

Authors:  Jürgen Wilbert; Matthias Guckenberger; Bülent Polat; Otto Sauer; Michael Vogele; Michael Flentje; Reinhart A Sweeney
Journal:  Radiat Oncol       Date:  2010-05-26       Impact factor: 3.481

5.  Frame and frameless linear accelerator-based radiosurgery for idiopathic trigeminal neuralgia.

Authors:  Allan Y Chen; Yen Hsieh; Steffanie McNair; Qijuan Li; Kevin Y Xu; Conrad Pappas
Journal:  J Radiosurg SBRT       Date:  2015

6.  Impact of PTV margin reduction (2 mm to 0 mm) on pseudoprogression in stereotactic radiotherapy of solitary brain metastases.

Authors:  Justine Badloe; Mirjam Mast; Anna Petoukhova; Jan-Huib Franssen; Elyas Ghariq; Noëlle van der Voort van Zijp; Ruud Wiggenraad
Journal:  Tech Innov Patient Support Radiat Oncol       Date:  2021-03-03

7.  Prolonged treatment time deteriorates positioning accuracy for stereotactic radiosurgery.

Authors:  Chun-Wei Wang; Yin-Chun Lin; Ham-Min Tseng; Furen Xiao; Chang-Mu Chen; Wei-Li Cheng; Szu-Huai Lu; Keng-Hsueh Lan; Wan-Yu Chen; Hsiang-Kuang Liang; Sung-Hsin Kuo
Journal:  PLoS One       Date:  2015-04-20       Impact factor: 3.240

8.  Evaluation of the setup accuracy of a stereotactic radiotherapy head immobilization mask system using kV on-board imaging.

Authors:  Imad Ali; Jesse Tubbs; Kerry Hibbitts; Ozer Algan; Spencer Thompson; Terence Herman; Salahuddin Ahmad
Journal:  J Appl Clin Med Phys       Date:  2010-05-20       Impact factor: 2.102

9.  Quantitative evaluation of patient setup uncertainty of stereotactic radiotherapy with the frameless 6D ExacTrac system using statistical modeling.

Authors:  Vance Keeling; Sabbir Hossain; Hosang Jin; Ozer Algan; Salahuddin Ahmad; Imad Ali
Journal:  J Appl Clin Med Phys       Date:  2016-05-08       Impact factor: 2.102

10.  Clinical Positioning Accuracy for Multisession Stereotactic Radiotherapy With the Gamma Knife Perfexion.

Authors:  Wade P Smith; Lori A Young; Mark H Phillips; Michael Cheung; Lia M Halasz; Jason K Rockhill
Journal:  Technol Cancer Res Treat       Date:  2017-05-17
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.