Unusual pattern of glistening formation on a 3-piece hydrophobic acrylic intraocular lens.

Liliana Werner1, Joachim Storsberg, Olivia Mauger, Karl Brasse, Ralph Gerl, Matthias Müller, Manfred Tetz.   

Abstract

A 3-piece hydrophobic acrylic intraocular lens (IOL) was explanted from the left eye of a 68-year-old patient 26 days after implantation because of glistenings that impaired fundus visualization. The effect of the glistenings on the patient's visual function was unknown because of retinal issues. The fluid-filled vacuoles within the IOL optic were different from those described in clinical and experimental studies. They were initially observed 4 days after surgery and were very large. The glistening formation was reproduced experimentally by immersing the explanted IOL in balanced salt solution at 37 degrees C and analyzing the IOL at room temperature for up to 2 weeks. An IOL of the same design was used as a control. Although both IOLs were subjected to the same laboratory conditions, gross and light microscopic analyses confirmed that the pattern of the glistening formation differed between them. Analyses of the explanted IOL and the control IOL under differential scanning calorimetry, as well as by attenuated total reflection Fourier transform infrared spectroscopy, revealed slight differences between the IOLs.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18721729     DOI: 10.1016/j.jcrs.2008.04.048

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  7 in total

1.  Decreased visual function due to high-level light scattering in a hydrophobic acrylic intraocular lens.

Authors:  Shinichiro Yoshida; Hiroyuki Matsushima; Mayumi Nagata; Tadashi Senoo; Ichiro Ota; Kensaku Miyake
Journal:  Jpn J Ophthalmol       Date:  2011-02-18       Impact factor: 2.447

2.  Analysis of glistenings in hydrophobic acrylic intraocular lenses on visual performance.

Authors:  Lei Xi; Yi Liu; Feng Zhao; Chuan Chen; Bing Cheng
Journal:  Int J Ophthalmol       Date:  2014-06-18       Impact factor: 1.779

Review 3.  Nanomaterials-Tools, Technology and Methodology of Nanotechnology Based Biomedical Systems for Diagnostics and Therapy.

Authors:  Christian Schmidt; Joachim Storsberg
Journal:  Biomedicines       Date:  2015-07-20

4.  Laboratory analyses of two explanted hydrophobic acrylic intraocular lenses.

Authors:  Yunhai Dai; Yusen Huang; Ting Liu; Lixin Xie
Journal:  Indian J Ophthalmol       Date:  2014-06       Impact factor: 1.848

5.  Decreased visual acuity resulting from glistening and sub-surface nano-glistening formation in intraocular lenses: A retrospective analysis of 5 cases.

Authors:  Hiroyuki Matsushima; Mayumi Nagata; Yoko Katsuki; Ichiro Ota; Kensaku Miyake; George H H Beiko; Andrzej Grzybowski
Journal:  Saudi J Ophthalmol       Date:  2015-07-29

6.  Comparison of glistenings formation and their effect on forward light scatter between the Acrysof SN60WF and Eternity Natural Uni NW-60 intraocular lenses.

Authors:  Eva Philippaki; David P O'Brart; Chistopher C Hull
Journal:  BMJ Open Ophthalmol       Date:  2020-02-12

7.  The assessment of the impact of glistening on visual performance in relation to tear film quality.

Authors:  Amanda Argay; Peter Vamosi
Journal:  PLoS One       Date:  2020-10-12       Impact factor: 3.240

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.