Validation of the SoloSTAR insulin pen.

Sherwyn Schwartz1, Aleksandra Vlajnic.   

Abstract

BACKGROUND: This study was designed to validate the use of SoloSTAR (sanofi-aventis Deutschland GmbH, Frankfurt, Germany), a new disposable pen developed for use with insulin glargine by patients with type 1 or type 2 diabetes.
METHODS: This single-center, open-label, single-arm, sequential trial enrolled subjects with type 1 or type 2 diabetes, 21-78 years old. After face-to-face training (Part 1) or self-training (Part 2), subjects performed three dose-delivery repetitions into an injection pad using separate pens; pens were weighed before and after each dose delivery. The primary outcome was the proportion of subjects delivering successful doses (target dose, 40 units) with all three repetitions. Secondary outcomes included pen accuracy and precision.
RESULTS: Validation population included 50 (Part 1) and 54 (Part 2) subjects. In Part 1, 100% of the subjects delivered a successful dose on all three repetitions. In Part 2, 98% of subjects delivered a successful dose on all three repetitions, with five cases of dose-delivery failures reported in the study population. The mean dose delivered was 40.2 units (95% confidence interval [CI], 40.1-40.3 units) in Part 1 and 38.0 units (95% CI, 36.7-39.3 units) in Part 2; 99% and 88% of dose repetitions in Parts 1 and 2, respectively, were within 38-42 units. The safety test was successfully performed by 80% (95% CI, 70.2-88.0%) of subjects in Part 1 and 68% (95% CI, 58.1-77.6%) in Part 2.
CONCLUSIONS: This study successfully validated the SoloSTAR pen for use by subjects with or without face-to-face training. SoloSTAR was shown to be easy to use, accurate, and precise.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18715210     DOI: 10.1089/dia.2008.0079

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  2 in total

1.  A comparison of dosing accuracy: visually impaired and sighted people using insulin pens.

Authors:  Ann S Williams; Patrick A Schnarrenberger
Journal:  J Diabetes Sci Technol       Date:  2010-05-01

2.  User-Centric Approach to Specifying Technical Attributes of Drug Delivery Devices: Empirical Study of Autoinjector-Cap Removal Forces.

Authors:  Andreas Schneider; Philipp Richard; Philippe Mueller; Christoph Jordi; Mary Yovanoff; Jakob Lange
Journal:  Patient Prefer Adherence       Date:  2021-02-02       Impact factor: 2.711

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.