Delayed rupture of flexor tendons caused by protrusion of a screw head of a volar plate for distal radius fracture: a case report.

Hiroshi Yamazaki1, Yasunori Hattori, Kazuteru Doi.   

Abstract

We report a rare occurrence of attritional rupture of flexor tendons caused by protrusion of the screw head of the volar radius plate. The cause of the protrusion was plate placement on the prominent distal volar lip of the radius and secondary loss of the original reduction due to using a non-locking osteosynthesis system for the treatment of displaced intra-articular fracture of the distal radius.

Entities:  

Mesh:

Year:  2008        PMID: 18711781     DOI: 10.1142/S0218810408003785

Source DB:  PubMed          Journal:  Hand Surg        ISSN: 0218-8104


  4 in total

1.  The cadaveric anatomy of the distal radius: implications for the use of volar plates.

Authors:  P A McCann; D Clarke; R Amirfeyz; R Bhatia
Journal:  Ann R Coll Surg Engl       Date:  2012-03       Impact factor: 1.891

2.  Rupture of Flexor Pollicis Longus Tendon: A Complication of Volar Locking Plating of the Distal Radius.

Authors:  Aysha Sethunathan Rajeev; Shanaka Sreverthana; John Harrison
Journal:  Eur J Trauma Emerg Surg       Date:  2009-09-11       Impact factor: 3.693

3.  Titanium versus Stainless-Steel Plating in the Surgical Treatment of Distal Radius Fractures: A Randomized Trial.

Authors:  Sameer Shakir; Sanjay Naran; Mithun Neral; Ronit Wollstein
Journal:  J Hand Microsurg       Date:  2016-10-12

Review 4.  Flexor tendon injuries following plate fixation of distal radius fractures: a systematic review of the literature.

Authors:  Saeed Asadollahi; Prue P A Keith
Journal:  J Orthop Traumatol       Date:  2013-05-14