Literature DB >> 18711244

Thimble ionization chambers in medium-energy x-ray beams and the role of constructive details of the central electrode: Monte Carlo simulations and measurements.

F Ubrich1, J Wulff, R Kranzer, K Zink.   

Abstract

This paper presents investigations of thimble ionization chamber response in medium-energy kilovoltage x-ray beams (70-280 kVp, 0.09-3.40 mm Cu HVL). Two thimble ionization chambers (PTW30015 and PTW30016) were investigated, regarding the influence of the central electrode dimensions made of aluminum. Measurements were carried out in photon fields of different beam quality. Corresponding Monte Carlo simulations employing the EGSnrc Monte Carlo code system were performed. The simulations included the modelling of the x-ray tube and measurement setup for generation of x-ray spectra. These spectra were subsequently used to calculate the absorbed energy in the air cavity of the two thimble ionization chamber models and the air kerma at the reference point of the chambers. Measurements and simulations revealed an optimal diameter of the central electrode, concerning an almost energy-independent response of the ionizaton chamber. The Monte Carlo simulations are in good agreement with the measured values, expressed in beam quality correction factors k(Q). The agreement was generally within 0.6% but could only be achieved with an accurate model of the central electrode including its exact shape. Otherwise, deviations up to 8.5% resulted, decreasing with higher photon energies, which can be addressed to the high yield of the photoelectric effect in the electrode material aluminum at low photon energies.

Entities:  

Mesh:

Year:  2008        PMID: 18711244     DOI: 10.1088/0031-9155/53/18/003

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  1 in total

1.  Enhanced Tomographic Sensing Multimodality with a Crystal Analyzer.

Authors:  Alexey Buzmakov; Marina Chukalina; Irina Dyachkova; Anastasia Ingacheva; Dmitry Nikolaev; Denis Zolotov; Igor Schelokov
Journal:  Sensors (Basel)       Date:  2020-12-06       Impact factor: 3.576

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.