[Effect of tear film changes on the intraocular pressure measurement by Goldmann applanation tonometer].

Yangfa Zeng1, Xinxing Guo, Jun Lin, Xiangyun Zeng, Yiming Zhong, Xiaoyu Cai, Xing Liu.   

Abstract

PURPOSE: To explore the effect of tear film changes on the intraocular pressure (IOP) measurement by Goldmann applanation tonometer.
METHODS: Sixty-eight normal subjects were randomly divided into two groups (Group A and B), 34 in each group. One eye of each subject by random selection was given Dextran (in Group A) and Viscoat (in Group B), respectively. And the fellow eyes were used as controls. IOP was measured in all subjects twice using Goldmann applanation tonometer. And the difference between first and second measurements was compared. RESULT: In the control eyes of Group A, the average IOP of first and second measurements were (14.44 +/- 2.68) mmHg (1 mmHg = 0.133 KPa) and (14.47 +/- 2.69) mmHg, there was no difference between the two measurements (t = -0.329, P = 0.744). In eyes given Dextran, the average IOP of two measurements were (14.41 +/- 2.63) mmHg and (12.94 +/- 2.59) mmHg, there was significant difference between the two measurements (t = 13.949, P = 0.000). In control eyes of Group B, the average IOP of first and second measurements were (14.53 +/- 3.13) mmHg and (11.18 +/- 3.07) mmHg, and there was no difference between the two measurements (t = -0.466, P = 0.644). In eyes given Viscoat, the average IOP of two measurements were (14.53 +/- 3.13) mmHg and (11.18 +/- 3.07) mmHg, and there was significant difference between the two measurements (t = 22.126, P = 0.000). After being given Dextran or Viscoat, the IOP values decreased consistently in both Group A and Group B.
CONCLUSIONS: The change of tear film components can affect IOP values by Goldmann applanation tonometer.

Entities:  

Mesh:

Year:  2008        PMID: 18709953

Source DB:  PubMed          Journal:  Yan Ke Xue Bao        ISSN: 1000-4432


  2 in total

1.  Goldmann tonometry tear film error and partial correction with a shaped applanation surface.

Authors:  Sean J McCafferty; Eniko T Enikov; Jim Schwiegerling; Sean M Ashley
Journal:  Clin Ophthalmol       Date:  2018-01-04

2.  The influence of the tonometer position on canine intraocular pressure measurements using the Tonovet® rebound tonometer.

Authors:  Juliana Kravetz de Oliveira; Fabiano Montiani-Ferreira; David L Williams
Journal:  Open Vet J       Date:  2018-03-09
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.