| Literature DB >> 18699345 |
G D Wedlake1, J H Vera, G A Ratcliff.
Abstract
A new analysis of the factors affecting the driving head within a capillary viscometer has identified a previously undetected source of error due to surface tension. In contrast to the well-known capillary rise correction to the effective driving head, this additional effect requires that a correction be made to the apparent head of liquid. Calculations show that this source of error may be responsible for deviations in experimental results as large as 1%. A new calibration equation which compensates for both surface tension effects is recommended.Year: 1979 PMID: 18699345 DOI: 10.1063/1.1135642
Source DB: PubMed Journal: Rev Sci Instrum ISSN: 0034-6748 Impact factor: 1.523