Literature DB >> 18699312

Apparatus for the measurement of small absorption change kinetics at 820 nm in the nanosecond range after a ruby laser flash.

J A Van Best1, P Mathis.   

Abstract

An apparatus for the measurement of kinetics of small absorption changes (DeltaA approximately 10(-3)-10(-4)) at 820 nm in a biological sample in the nanosecond range after a short high-intensity ruby laser flash (200 mJ, 8 ns) is described. The laser-induced fluorescence artefact is greatly reduced by the use of a solid-state laser diode as source of measuring light, in combination with a microscope lens and a diaphragm. The signal-to-noise ratio is enhanced by the use of an avalanche photodetector connected to a laboratory-made low-noise and interference-free amplifier. The signals are processed by a transient digitizer in combination with a signal averager. The response time of the apparatus is about 20 ns. With some modifications the apparatus can be used to measure absorption changes in a broad spectral range.

Year:  1978        PMID: 18699312     DOI: 10.1063/1.1135579

Source DB:  PubMed          Journal:  Rev Sci Instrum        ISSN: 0034-6748            Impact factor:   1.523


  2 in total

1.  Spectral and kinetic characterization of electron acceptor A1 in a Photosystem I core devoid of iron-sulfur centers F X, F B and F A.

Authors:  K Brettel; J H Golbeck
Journal:  Photosynth Res       Date:  1995-09       Impact factor: 3.573

2.  Pushing the limits of flash photolysis to unravel the secrets of biological electron and proton transfer.

Authors:  Paul Mathis; Evelyne Sage; Martin Byrdin
Journal:  Photochem Photobiol Sci       Date:  2022-01-26       Impact factor: 4.328

  2 in total