Bovine bone screws: metrology and effects of chemical processing and ethylene oxide sterilization on bone surface and mechanical properties.

D P Haje1, J B Volpon, C A Moro.   

Abstract

We assess the effects of chemical processing, ethylene oxide sterilization, and threading on bone surface and mechanical properties of bovine undecalcified bone screws. In addition, we evaluate the possibility of manufacturing bone screws with predefined dimensions. Scanning electronic microscopic images show that chemical processing and ethylene oxide treatment causes collagen fiber amalgamation on the bone surface. Processed screws hold higher ultimate loads under bending and torsion than the in natura bone group, with no change in pull-out strength between groups. Threading significantly reduces deformation and bone strength under torsion. Metrological data demonstrate the possibility of manufacturing bone screws with standardized dimensions.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18697880     DOI: 10.1177/0885328208092110

Source DB:  PubMed          Journal:  J Biomater Appl        ISSN: 0885-3282            Impact factor:   2.646


  1 in total

1.  EFFECTS OF CHEMICAL PROCESSING AND OXIDE ETHYLENE STERILIZATION ON CORTICAL AND CANCELLOUS RAT BONE: A LIGHT AND ELECTRON SCANNING MICROSCOPY STUDY.

Authors:  Marcello Teixeira Castiglia; Juliano Voltarelli F da Silva; José Armendir Frezarim Thomazini; José Batista Volpon