The comparison of doses measured by radiochromic films and semiconductor detector in a 175MeV proton beam.

Marta Mumot1, Gennady V Mytsin, Alexander G Molokanov, Julian Malicki.   

Abstract

We wanted to verify the response of radiochromic films in a 175MeV clinical proton beam used at the Joint Institute for Nuclear Research in Dubna against doses measured using semiconductor detectors and compare the results with published data from other centres. Radiochromic films (RCFs) MD-55 and a Vidar VXR-16 scanner were used. The films were irradiated in an unmodulated proton beam and with a beam modulated with a bolus and a ridge filter. Obtained dose distributions were compared with dose distributions measured with a Si-semiconductor detector. For the unmodulated beam the difference between the RCF and the semiconductor detector was 12% in the Bragg peak top. For the modulated beam the difference inside the spread-out Bragg peak region was 4%. Observed deviations between doses measured with RCF and Si-detector outside the Bragg peak were caused by the inhomogeneity of radiochromic emulsion. In the Bragg peak region the RCF doses were lower than those measured by semiconductors. The results were in agreement with published data from other proton therapy centres.

Mesh:

Substances:

Year:  2008        PMID: 18694653     DOI: 10.1016/j.ejmp.2008.06.001

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  1 in total

1.  Wax boluses and accuracy of EBT and RTQA radiochromic film detectors in radiotherapy with the JINR Phasotron proton beam.

Authors:  Dorota Maria Borowicz; Julian Malicki; Gennady Mytsin; Konstantin Shipulin
Journal:  Rep Pract Oncol Radiother       Date:  2013-07-02
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.