Mélissa Jourdain1, Jean Meunier, Jean Sequeira, Jean-Marc Boï, Jean-Claude Tardif.
Abstract
During an intravascular ultrasound (IVUS) intervention, a catheter with an ultrasound transducer is introduced in the body through a blood vessel, and then, pulled back to image a sequence of vessel cross sections. Unfortunately, there is no 3-D information about the position and orientation of these cross-section planes, which makes them less informative. To position the IVUS images in space, some researchers have proposed complex stereoscopic procedures relying on biplane angiography to get two X-ray image sequences of the IVUS transducer trajectory along the catheter. To simplify this procedure, we and others have elaborated algorithms to recover the transducer 3-D trajectory with only a single view X-ray image sequence. In this paper, we present an improved method that provides both automated 2-D and 3-D transducer tracking based on pullback speed as a priori information. The proposed algorithm is robust to erratic pullback speed and is more accurate than the previous single-plane 3-D tracking methods.Mesh:
Year: 2008 PMID: 18693498 DOI: 10.1109/titb.2008.921043
Source DB: PubMed Journal: IEEE Trans Inf Technol Biomed ISSN: 1089-7771