| Literature DB >> 18693306 |
Laura Mercolini1, Roberto Mandrioli, Mario Amore, Maria Augusta Raggi.
Abstract
Benzodiazepines (BZDs) are often prescribed to schizophrenic or depressed patients, as a part of polypharmacy regimens. An HPLC method has been developed for the simultaneous determination of 15 BZDs in human plasma. Separation was obtained by using a C8 RP column and a mobile phase composed of 65% aqueous phosphate buffer at pH 3.0 and 35% ACN. The UV detector was set at 220 nm and clomipramine was used as the internal standard. A careful pretreatment procedure of plasma samples was developed, using SPE with C1 cartridges, which gives high extraction yields (> 97%). The LOQs were always lower than 7.6 ng/mL and the LODs always lower than 2.6 ng/mL for all analytes. The method was successfully applied to plasma samples from depressed and schizophrenic patients undergoing polypharmacy with one or more BZDs. Precision data, as well as accuracy results, were satisfactory and no interference from other drugs was found. Hence, the method seems to be suitable for the therapeutic drug monitoring (TDM) of patients undergoing therapy with one or more BZDs.Entities:
Mesh:
Substances:
Year: 2008 PMID: 18693306 DOI: 10.1002/jssc.200800212
Source DB: PubMed Journal: J Sep Sci ISSN: 1615-9306 Impact factor: 3.645