Literature DB >> 18687844

A report on the early failure of the LPM proximal interphalangeal joint replacement.

J L Hobby1, S Edwards, J Field, G Giddins.   

Abstract

The LPM (Van Straten Medical, Netherlands) proximal interphalangeal joint replacement was introduced to the United Kingdom in October 2000, and 257 implants were sold. We have audited the results of this prosthesis, and obtained data for 164 of the 257 implants. Forty-seven (29%) of the 164 have been revised and another 33 (20%) are showing clinical and radiological signs of failure at a maximum follow-up of 6 years. Massive osteolysis leading to aseptic loosening is the commonest cause of failure. We believe that all patients who have had this prosthesis inserted should be kept under regular clinical and radiographic review.

Entities:  

Mesh:

Year:  2008        PMID: 18687844     DOI: 10.1177/1753193408094920

Source DB:  PubMed          Journal:  J Hand Surg Eur Vol        ISSN: 0266-7681


  3 in total

1.  Wear testing of a DJOA finger prosthesis in vitro.

Authors:  Thomas J Joyce
Journal:  J Mater Sci Mater Med       Date:  2010-02-11       Impact factor: 3.896

2.  Surface replacement arthroplasty of the proximal interphalangeal and metacarpophalangeal joints: The current state.

Authors:  Harvinder Singh; Joseph J Dias
Journal:  Indian J Plast Surg       Date:  2011-05

3.  New implants in orthopaedics--caveat emptor.

Authors:  D Warwick; I A Trail
Journal:  Ann R Coll Surg Engl       Date:  2012-10       Impact factor: 1.891

  3 in total