Effects of continuous and triggered atrial overdrive pacing on paroxysmal atrial fibrillation in pacemaker patients.

Andreas Schuchert1, Hans-Peter Rebeski, Thomas Peiffer, Eberhard Bub, Armin Dietz, Kai Mortensen, Mohammed Ali Aydin, John Camm, Steffen Gazarek, Thomas Meinertz.   

Abstract

BACKGROUND: The aim of the study was to compare the effects of different pacing strategies to prevent paroxysmal atrial fibrillation (AF): triggered atrial overdrive pacing versus the combination of triggered and continuous overdrive pacing.
METHODS: Patients with an indication for dual-chamber pacing (Selection 9000, Prevent AF; Vitatron B.V., Arnhem, the Netherlands) and a history of paroxysmal AF were randomized to triggered atrial pacing (three pacing functions, "triggered group": PAC Suppression, Post-PAC Response, and Post-Exercise Response) or to the combination of continuous (Pace Conditioning) and triggered atrial pacing (four pacing functions, "combined group"). After 3 months, there was a crossover to the other pacemaker setting.
RESULTS: In 171 enrolled patients, the median AF burden of the combined group was with 2.1% versus 0.1% in the triggered group (P = 0.014). Fewer AF episodes were observed in the triggered (median: 7) than in the combined group (median: 116; P = 0.016). The combined group had more frequent atrial pacing (median 97%) than the triggered group with 85% (P < 0.001), but ventricular pacing was not significantly different with 95% and 96% in the combined and triggered group, respectively. After the crossover, the AF burden increased in the triggered group to 0.3% and decreased in the combined group to 0.4%.
CONCLUSIONS: Triggered atrial pacing functions alone resulted in a low AF burden. The additional activation of continuous atrial overdrive pacing increased the percentage of atrial pacing, but had no beneficial effects on the prevention of paroxysmal AF.

Entities:  

Mesh:

Year:  2008        PMID: 18684247     DOI: 10.1111/j.1540-8159.2008.01118.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  2 in total

1.  [Atrial sensing with fixed and floating electrodes at identical activities].

Authors:  D Hartung; L Thiele; W M Hartung
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2009-04-18

2.  [Atrial pacing for prevention of atrial fibrillation. Influence of septal atrial pacing, atrial overdrive and AV-delay-optimization on atrial fibrillation burden].

Authors:  M Klein; N Klein; D Pfeiffer
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2009-09
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.