| Literature DB >> 18681702 |
L Aguilera1, M Lanza, M Porti, J Grifoll, M Nafría, X Aymerich.
Abstract
A new configuration of conductive atomic force microscope (CAFM) is presented, which is based in a standard CAFM where the typical I-V converter has been replaced by a log I-V amplifier. This substitution extends the current dynamic range from 1-100 pA to 1 pA-1 mA. With the broadening of the current dynamic range, the CAFM can access new applications, such as the reliability evaluation of metal-oxide-semiconductor gate dielectrics. As an example, the setup has been tested by analyzing breakdown spots induced in SiO2 layers.Entities:
Year: 2008 PMID: 18681702 DOI: 10.1063/1.2952058
Source DB: PubMed Journal: Rev Sci Instrum ISSN: 0034-6748 Impact factor: 1.523