Temperature cycler evaluation: what do you need to know?

G P Tweed1, J Whitney, P L Bloch.   

Abstract

Which of the 20 different, commercially available, temperature cyclers adequately fulfill all functional requirements? How can an investigator analyze instrument function? This report describes in detail, for the first time, the characteristics of Ericomp temperature cyclers. The data show that the temperature uniformity, accuracy and maximum overshoot is +/- 0.34 degree C, less than 0.5 degree C and 0.17 degree C, respectively. Cycling temperature reproducibility is 0.1%, and cycling time reproducibility is 0.4%. These data compare very favourably to published data describing other temperature cyclers. The techniques used to obtain these data serve as a model for evaluating instruments for well-to-well uniformity, temperature accuracy and within-run reproducibility.

Mesh:

Substances:

Year:  1991        PMID: 1867863

Source DB:  PubMed          Journal:  Biotechniques        ISSN: 0736-6205            Impact factor:   1.993


  1 in total

1.  Novel approach for assessing performance of PCR cyclers used for diagnostic testing.

Authors:  D Schoder; A Schmalwieser; G Schauberger; J Hoorfar; M Kuhn; M Wagner
Journal:  J Clin Microbiol       Date:  2005-06       Impact factor: 5.948