Abstract

BACKGROUND: Despite considerable recent interest in computer navigation for orthopaedic surgery, few investigations of computer-assisted surgery for foot and ankle operations have been reported. The purpose of the present study was to compare subtalar arthrodesis with and without computer navigation in a cadaver model.
METHODS: Subtalar arthrodesis was performed on thirty-six matched-pair cadaver lower extremities with intact soft tissues, with an attempt being made to orient two screws in the optimal configuration based on unpublished data from a preceding biomechanical study. Each matched pair was randomly assigned either to a group of surgeons who were experienced in subtalar arthrodesis or to a group of inexperienced operators. Neither surgical group was experienced in computer-assisted surgery. We compared optimal first-pass guidewire placement, fluoroscopic time, total operative time, screw placement accuracy, and adverse screw placement events between conventional (fluoroscopically guided) and computer-assisted subtalar arthrodesis.
RESULTS: The number of passes needed to achieve optimal guidewire placement decreased with the use of computer assistance for both experienced surgeons and inexperienced operators (p < 0.001), with ideal placement occurring on the first attempt in 95% of the procedures performed with use of computer assistance. While the experienced surgeons required less time and fewer guidewire passes during conventional subtalar arthrodesis than the inexperienced operators did (p < 0.001), both groups used less fluoroscopy with computer assistance (p < 0.001). There was no significant difference in operative time between the two techniques when performed by the inexperienced operators, yet the total procedure time doubled for the experienced surgeons when the procedure was performed with use of computer assistance (p < 0.001). There was no significant difference between experienced surgeons and inexperienced operators or between conventional and computer-assisted subtalar arthrodesis with respect to adverse screw placement events or the ability to accurately place both screws.
CONCLUSIONS: Computer-assisted subtalar arthrodesis resulted in screw placement accuracy that was equivalent to that of conventional (fluoroscopically guided) subtalar arthrodesis while decreasing the number of suboptimal guidewire passes and fluoroscopic time. The computer-assisted surgery technique increased the operative time for surgeons who were more experienced in conventional subtalar arthrodesis, but there was no difference in operative time for the group of operators who were inexperienced in subtalar arthrodesis.

Mesh:

Year:  2008        PMID: 18676891     DOI: 10.2106/JBJS.G.00513

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  3 in total

1.  Intraoperative navigation for minimally invasive resection of periarticular and pelvic tumors.

Authors:  Kevin Wu; Nicholas P Webber; Russell A Ward; Kevin B Jones; R Lor Randall
Journal:  Orthopedics       Date:  2011-05       Impact factor: 1.390

2.  MRI-guided navigation surgery with temporary implantable bone markers in limb salvage for sarcoma.

Authors:  June Hyuk Kim; Hyun Guy Kang; Han-Soo Kim
Journal:  Clin Orthop Relat Res       Date:  2010-01-07       Impact factor: 4.176

3.  Minimally invasive, no hardware subtalar arthrodesis with autogenous posterior iliac bone graft.

Authors:  Mohamed A M Eid; Maged Abou El-Soud; Mahmoud A Mahran; Timour F El-Hussieni
Journal:  Strategies Trauma Limb Reconstr       Date:  2010-03-04
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.