Literature DB >> 18670852

Improvement of the diagnostic accuracy of lymph node metastases of colorectal cancer in 18F-FDG-PET/CT by optimizing the iteration number for the image reconstruction.

Kazumasa Inoue1, Takashi Sato, Hideaki Kitamura, Masaaki Ito, Yoshiyuki Tsunoda, Akira Hirayama, Hideo Kurosawa, Takashi Tanaka, Masahiro Fukushi, Noriyuki Moriyama, Hirofumi Fujii.   

Abstract

OBJECTIVE: Improvement in the diagnostic accuracy of lymph node (LN) metastases of colorectal cancer by optimizing the reconstruction parameters was studied, including a phantom study and clinical studies.
METHODS: In the experimental study, the contrast ratio was evaluated using a standard image quality phantom, changing the iteration number of ordered subsets expectation maximization algorithm from 2 to 6. In the clinical study, 89 patients with preoperative colorectal cancer who received 18F-2-deoxy-2-fluoro-D: -glucose positron emission tomography/computed tomography (PET/CT) were studied. Their LN metastases were visually assessed when systematically changing the iteration number, and the optimal iteration number was determined. After the appropriate cut-off value of maximum standardized uptake value was determined, the improvement of the diagnostic accuracy of LN metastases was evaluated for the proximal nodes and the distal ones using the reconstructed images with the optimal iteration number. This was compared with the conventional method that had an iteration number of 2.
RESULTS: In the phantom study, it was confirmed that the contrast ratio improved when the iteration number increased. In clinical evaluation, the optimal iteration number was determined to be 5 by visual assessment. When the cut-off value of 1.5 was used, which happened to be the best number, the sensitivity/specificity/accuracy improved from 52%/91%/72% to 77%/89%/83% for the proximal nodes, and from 63%/90%/84% to 91%85%/87% for the distal nodes, respectively.
CONCLUSIONS: When the iteration number of the reconstruction algorithm was optimized, the sensitivity of LN metastasis improved by more than 20%, and the accuracy exceeded 80%. Optimization of the image reconstruction parameters in the diagnosis of LN metastases using PET/CT is clinically important.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18670852     DOI: 10.1007/s12149-007-0141-9

Source DB:  PubMed          Journal:  Ann Nucl Med        ISSN: 0914-7187            Impact factor:   2.668


  2 in total

1.  Segmental acquisition method for stationary objects in (18)F-fluorodeoxyglucose positron emission tomography tests.

Authors:  Keisuke Tsuda; Naoyuki Aikawa; Takayuki Suzuki; Etsuo Moriya; Masayuki Yamaguchi; Hideaki Kitamura; Kouzou Hanai; Izumi O Umeda; Masahiro Fukushi; Noriyuki Moriyama; Hirofumi Fujii
Journal:  Jpn J Radiol       Date:  2010-10-24       Impact factor: 2.374

2.  Diagnostic value of FDG-PET/CT for lymph node metastasis of colorectal cancer.

Authors:  Jae Young Kwak; Jae Seung Kim; Hye Jin Kim; Hyun Kwon Ha; Chang Sik Yu; Jin Cheon Kim
Journal:  World J Surg       Date:  2012-08       Impact factor: 3.352

  2 in total