Intrafraction changes of prostate position and geometrical errors studied by continuous electronic portal imaging.

Torbjørn Månsson Haskå1, Henriette Honore, Ludvig Paul Muren, Morten Høyer, Per Rugaard Poulsen.   

Abstract

PURPOSE: The use of marker-based on-line image guided radiotherapy for prostate cancer has considerably reduced the treatment margins to sub-cm. In this study we have quantified the residual set-up errors remaining after isocenter correction, studied their development during beam delivery and estimated their impact on margins. METHODS AND MATERIALS: After initial on-line patient set-up based on orthogonal kV x-ray images of implanted fiducial markers, continuous electronic portal imaging was performed during treatment delivery in 10 of 39 treatment sessions for 20 prostate cancer patients. The cranio-caudal (CC) position of the centre-of-mass of the three markers was found using a threshold technique on every single image frame for all patients, typically 12-14 images for 5 treatment beams in every fraction. The CC prostate position was determined relative to its initial position at treatment onset and relative to its planned position within the field aperture. These results allowed determination of the CC intrafraction prostate motion and the intrafraction progression of the geometrical CC error, respectively.
RESULTS: At treatment onset the standard deviation (SD) of the set-up error was 1.0mm in the lateral direction and 1.5mm in the cranio-caudal (CC) direction. It did not depend significantly on the duration of the set-up procedure (mean: 3.0 min, span 1.2-14.6 min). The distribution of CC prostate positions relative to the position at treatment onset broadened from 0 to 1.4mm during the treatment session, while the corresponding CC setup error distribution broadened from 1.5 to 1.9 mm. This broadening means that the necessary CC setup margin increased by around 1mm during the treatment fraction.
CONCLUSIONS: Large differences in the intrafraction CC prostate motion patterns were found, however, intrafraction motion only results in a modest additional CC set-up margin of around 1mm relative to the margins needed for the residual set-up error at treatment start.

Entities:  

Mesh:

Year:  2008        PMID: 18663646     DOI: 10.1080/02841860802256509

Source DB:  PubMed          Journal:  Acta Oncol        ISSN: 0284-186X            Impact factor:   4.089


  3 in total

Review 1.  Target margins in radiotherapy of prostate cancer.

Authors:  Slav Yartsev; Glenn Bauman
Journal:  Br J Radiol       Date:  2016-07-20       Impact factor: 3.039

2.  Validation of planning target volume margins by analyzing intrafractional localization errors for 14 prostate cancer patients based on three-dimensional cross-correlation between the prostate images of planning CT and intrafraction cone-beam CT during volumetric modulated arc therapy.

Authors:  Kenshiro Shiraishi; Masahiko Futaguchi; Akihiro Haga; Akira Sakumi; Katsutake Sasaki; Kentaro Yamamoto; Hiroshi Igaki; Kuni Ohtomo; Kiyoshi Yoda; Keiichi Nakagawa
Journal:  Biomed Res Int       Date:  2014-05-22       Impact factor: 3.411

3.  Determining leaf trajectories for dynamic multileaf collimators with consideration of marker visibility: an algorithm study.

Authors:  Bo Zhao; Jianrong Dai
Journal:  J Radiat Res       Date:  2014-06-08       Impact factor: 2.724

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.