Operation of bolus tracking system for prediction of aortic peak enhancement at multidetector row computed tomography: pharmacokinetic analysis and clinical study.

Isao Yamaguchi1, Hiroyuki Hayashi, Masayuki Suzuki, Katsuhiro Ichikawa, Eiji Kidoya, Hirohiko Kimura.   

Abstract

PURPOSE: The present study was performed to identify the theoretical background for optimal use of the bolus tracking system by analyzing the changes in the initial slope of the aortic time-enhancement curve (TEC).
MATERIALS AND METHODS: We calculated the contrast medium aortic arrival time (TAR), the time to reach the trigger threshold (effective TAR), the slope of the linear equation of the enhancement unit (enhancement rate), and the time to peak aortic enhancement from the TECs of the pharmacokinetic analysis and retrospective clinical study.
RESULTS: In the pharmacokinetic analysis, the enhancement rate-simulated under conditions of injection duration 30 s and iodine load per body weight 500 mg/kg-was 27.1 HU/s. In the clinical study, the enhancement rate was 27.9 +/- 3.0 HU/s. A correlation was found between the TAR and the enhancement rate, indicating that enhancement rates decrease with increasing TAR. It took 22.7 +/- 0.5 s to reach maximum enhancement of the aorta from the trigger threshold of an increase of 100 HU and injection duration at 30 s.
CONCLUSION: We found that cardiac output differences are strongly dependent on the TAR and that most of the differences disappeared during the phase until effective TAR.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18661212     DOI: 10.1007/s11604-008-0228-9

Source DB:  PubMed          Journal:  Radiat Med        ISSN: 0288-2043


  27 in total

1.  Automatic bolus tracking in monophasic spiral CT of the liver: liver-to-lesion conspicuity.

Authors:  F Mehnert; P L Pereira; J Trübenbach; A F Kopp; C D Claussen
Journal:  Eur Radiol       Date:  2001       Impact factor: 5.315

2.  Peak contrast enhancement in CT and MR angiography: when does it occur and why? Pharmacokinetic study in a porcine model.

Authors:  Kyongtae T Bae
Journal:  Radiology       Date:  2003-04-17       Impact factor: 11.105

3.  Optimizing liver contrast in helical liver CT: value of a real-time bolus-triggering technique.

Authors:  H P Dinkel; M Fieger; J Knüpffer; R Moll; G Schindler
Journal:  Eur Radiol       Date:  1998       Impact factor: 5.315

4.  Dual-phase helical CT of the liver: effects of bolus tracking and different volumes of contrast material.

Authors:  L Kopka; J Rodenwaldt; U Fischer; D W Mueller; J W Oestmann; E Grabbe
Journal:  Radiology       Date:  1996-11       Impact factor: 11.105

5.  Contrast-enhanced CT with saline flush technique using two automated injectors: how much contrast medium does it save?

Authors:  Toshiyuki Irie; Motonao Kajitani; Masayuki Yamaguchi; Yuji Itai
Journal:  J Comput Assist Tomogr       Date:  2002 Mar-Apr       Impact factor: 1.826

6.  Small hypervascular hepatocellular carcinoma revealed by double arterial phase CT performed with single breath-hold scanning and automatic bolus tracking.

Authors:  Tonsok Kim; Takamichi Murakami; Masatoshi Hori; Manabu Takamura; Satoru Takahashi; Atsuya Okada; Shuji Kawata; Modesto Cruz; Michael P Federle; Hironobu Nakamura
Journal:  AJR Am J Roentgenol       Date:  2002-04       Impact factor: 3.959

7.  Aortic and hepatic enhancement and tumor-to-liver contrast: analysis of the effect of different concentrations of contrast material at multi-detector row helical CT.

Authors:  Kazuo Awai; Koichi Takada; Hiromitsu Onishi; Shinichi Hori
Journal:  Radiology       Date:  2002-09       Impact factor: 11.105

8.  Evaluation of optimal timing of arterial phase imaging for the detection of hypervascular hepatocellular carcinoma by using triple arterial phase imaging with multidetector-row helical computed tomography.

Authors:  Takamichi Murakami; Tonsok Kim; Syuji Kawata; Masayuki Kanematsu; Michael P Federle; Masatoshi Hori; Atsuya Okada; Seishi Kumano; Eiji Sugihara; Kaname Tomoda; Hironobu Nakamura
Journal:  Invest Radiol       Date:  2003-08       Impact factor: 6.016

9.  [Estimation of aortic time-enhancement curve in pharmacokinetic analysis: dynamic study by multi-detector row computed tomography].

Authors:  Isao Yamaguchi; Hiroyuki Hayashi; Masayuki Suzuki; Eiji Kidoya; Kyoji Higashimura
Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi       Date:  2007-06-20

10.  Effect of contrast material injection duration and rate on aortic peak time and peak enhancement at dynamic CT involving injection protocol with dose tailored to patient weight.

Authors:  Kazuo Awai; Kumiko Hiraishi; Shinichi Hori
Journal:  Radiology       Date:  2004-01       Impact factor: 11.105

View more
  1 in total

1.  Optimizing scan timing of hepatic arterial phase by physiologic pharmacokinetic analysis in bolus-tracking technique by multi-detector row computed tomography.

Authors:  Isao Yamaguchi; Eiji Kidoya; Masayuki Suzuki; Hirohiko Kimura
Journal:  Radiol Phys Technol       Date:  2010-09-25