Literature DB >> 18656682

Enhanced sensitivity for Cu(II) by a salicylidine-functionalized polysiloxane carbon paste electrode.

Hazem M Abu-Shawish1, Salman M Saadeh, Ahmad R Hussien.   

Abstract

A new approach for decreasing the detection limit for a copper(II) ion-selective electrode (ISE) is presented. The ISE is designed using salicylidine-functionalized polysiloxane in carbon paste. This work describes the attempts to develop the electrode and measurements of its characteristics. The new type of renewable three-dimensional chemically modified electrode could be used in a pH range of 2.3-5.4, and its detection limit is 2.7 x 10(-8)mol L(-1) (1.2 microg L(-1)). This sensor exhibits a good Nernstian slope of 29.4+/-0.5 mV/decade in a wide linear concentration range of 2.3 x 10(-7) to 1.0 x 10(-3)mol L(-1) of Cu(II). It has a short response time (approximately 8s) and noticeably high selectivity over other Cu(II) selective electrodes. Finally, it was satisfactorily used as an indicator electrode in complexometric titration with EDTA and determination of copper(II) in miscellaneous samples such as urine and various water samples.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18656682     DOI: 10.1016/j.talanta.2008.04.063

Source DB:  PubMed          Journal:  Talanta        ISSN: 0039-9140            Impact factor:   6.057


  1 in total

1.  Rapid determination of trace level copper in tea infusion samples by solid contact ion selective electrode.

Authors:  Aysenur Birinci; Hilal Eren; Fatih Coldur; Emine Coskun; Muberra Andac
Journal:  J Food Drug Anal       Date:  2016-04-12       Impact factor: 6.157