Emília Szájli1, Tamás Fehér, Katalin F Medzihradszky.
Abstract
MALDI-TOF MS has been applied by several groups to relative quantitative measurements. At the same time, the non-quantitative character of this method has been widely reported. We conducted experiments to test the reliability of this technique for quantitation using the statistical method of inverse confidence limit calculation for the first time in this context. The relationship between relative intensities of known amounts of standard peptides and their concentration ratios was investigated. We found that the concentration ratios determined by relative intensity measurements were highly inaccurate and strongly influenced by the molecular milieu of the sample analyzed. Thus, we emphasize the necessity of using the sample itself for calibration. We also performed experiments using an isotope-labeled derivative of the analyte as an internal standard for calibration line generation. As expected, the use of such standard led to a dramatic increase in precision and a less pronounced improvement in accuracy. We recommend performing a similar statistical analysis as a demonstration of reliability for every system where MALDI-TOF MS is used for quantitative measurements.Mesh:
Substances:
Year: 2008 PMID: 18653768 DOI: 10.1074/mcp.M800108-MCP200
Source DB: PubMed Journal: Mol Cell Proteomics ISSN: 1535-9476 Impact factor: 5.911