The effect of repeated applanation on subsequent IOP measurements.

Turki M AlMubrad1, Kelechi C Ogbuehi.   

Abstract

BACKGROUND: In studies aimed at assessing the accuracy and repeatability of non-contact tonometers, the order in which these tonometers and the Goldmann tonometer are used is usually randomised despite studies in the literature that demonstrate an ocular massage effect that occurs post-applanation but not after non-contact tonometry. The purpose of this study was to investigate the effect of repeated corneal applanation on subsequent assessments of IOP.
METHODS: Data were obtained from 65 left eyes of 65 young, oculovisual normals. Three sets of IOP measurements were obtained, one set with the Goldmann applanation tonometer and two with the Topcon CT80 non-contact tonometer (one set each before and after applanation with the Goldmann tonometer), in each one of two separate measurement sessions, one week apart.
RESULTS: The average (and SD) IOP measured with the Goldmann tonometer in the first session (14.8+/-2.9 mmHg) did not vary significantly from the IOP measured with the non-contact tonometer (pre-applanation) in both sessions or with the average Goldmann IOP in the second session. The bias (mean difference +/- SD) between methods was 0.3+/-1.4 mmHg and 0.4+/-1.4 mmHg, respectively, for the first and second sessions, with the CT80 (pre-applanation) recording the higher IOP in both sessions. The within-session repeatability coefficients were +/-2.3 mmHg, +/-2.6 mmHg, +/-2.1 mmHg and +/-2.0 mmHg for the CT80 (pre-applanation) in the first and second sessions, and the Goldmann tonometer in the first and second sessions, respectively. Test-retest repeatability coefficients were +/-2.8 mmHg and +/-2.5 mmHg for the CT80 (pre-applanation) and the Goldmann tonometer respectively. Post-applanation with the Goldmann tonometer, there was a statistically significant (p<0.05) reduction (1.5+/-1.2 mmHg in session 1) in the IOP measured with the non-contact tonometer in both sessions.
CONCLUSION: These results suggest that repeated corneal applanation leads to a statistically significant reduction in IOP on subsequent measurements.

Entities:  

Mesh:

Year:  2008        PMID: 18651843     DOI: 10.1111/j.1444-0938.2008.00298.x

Source DB:  PubMed          Journal:  Clin Exp Optom        ISSN: 0816-4622            Impact factor:   2.742


  9 in total

1.  The influence of soft contact lenses on the intraocular pressure measurement.

Authors:  P G Firat; C Cankaya; S Doganay; M Cavdar; S Duman; E Ozsoy; B Koc
Journal:  Eye (Lond)       Date:  2011-11-11       Impact factor: 3.775

2.  Using Deep Learning to Automate Goldmann Applanation Tonometry Readings.

Authors:  Ted Spaide; Yue Wu; Ryan T Yanagihara; Shu Feng; Omar Ghabra; Jonathan S Yi; Philip P Chen; Francy Moses; Aaron Y Lee; Joanne C Wen
Journal:  Ophthalmology       Date:  2020-04-25       Impact factor: 12.079

3.  Effect of corneal astigmatism on intraocular pressure measurement using ocular response analyzer and Goldmann applanation tonometer.

Authors:  Mana Hagishima; Kazutaka Kamiya; Fusako Fujimura; Tetsuya Morita; Nobuyuki Shoji; Kimiya Shimizu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2009-09-29       Impact factor: 3.117

4.  Data quality and clinical decision-making: do we trust machines blindly?

Authors:  Konrad Pesudovs; Raymond A Applegate
Journal:  Clin Exp Optom       Date:  2009-05       Impact factor: 2.742

5.  Performance of the PT100 noncontact tonometer in healthy eyes.

Authors:  Turki M AlMubrad
Journal:  Clin Ophthalmol       Date:  2011-05-19

6.  Two-position measurement of intraocular pressure by PT100 noncontact tonometry in comparison with Goldmann tonometry.

Authors:  Kelechi C Ogbuehi; John C Chijuka; Uchechukwu L Osuagwu
Journal:  Clin Ophthalmol       Date:  2011-09-06

7.  Age-Related Changes of Intraocular Pressure in Elderly People in Southern China: Lingtou Eye Cohort Study.

Authors:  Xiaotong Han; Yong Niu; Xinxing Guo; Yin Hu; William Yan; Mingguang He
Journal:  PLoS One       Date:  2016-03-17       Impact factor: 3.240

8.  Comparison of intraocular pressure as measured by three different non-contact tonometers and goldmann applanation tonometer for non-glaucomatous subjects.

Authors:  Seung Pil Bang; Chong Eun Lee; Yu Cheol Kim
Journal:  BMC Ophthalmol       Date:  2017-11-02       Impact factor: 2.209

9.  Comparison of three methods of tonometry in normal subjects: Goldmann applanation tonometer, non-contact airpuff tonometer, and Tono-Pen XL.

Authors:  Ihsan Yilmaz; Cigdem Altan; Ebru Demet Aygit; Cengiz Alagoz; Okkes Baz; Sibel Ahmet; Semih Urvasizoglu; Dilek Yasa; Ahmet Demirok
Journal:  Clin Ophthalmol       Date:  2014-06-07
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.