Literature DB >> 18651422

Irregular ventricular tachycardia underdetected by implantable cardioverter defibrillator device.

Athanasios G Manolis1, Spyros Kourouklis, Dimitrios Chatzis, Konstantinos Kouvelas, Zenon Kyriakides.   

Abstract

A case of sustained monomorphic ventricular tachycardia underdetected by a single chamber implantable cardioverter defibrillator because of RR interval irregularity is presented. The programmed stability criterion is responsible for the underdetection. Special attention must be paid when it comes to programming this detection parameter.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18651422

Source DB:  PubMed          Journal:  Cardiol J        ISSN: 1898-018X            Impact factor:   2.737


  1 in total

1.  A case of withheld defibrillator treatment for 9 minutes of ventricular tachycardia.

Authors:  Evan Grove; Vincent Y See; Timm M Dickfeld