A computational fluid dynamic study of stent graft remodeling after endovascular repair of thoracic aortic dissections.

Stephen W K Cheng1, Edward S K Lam, George S K Fung, Pei Ho, Albert C W Ting, Kwok W Chow.   

Abstract

OBJECTIVES: Significant stent graft remodeling commonly occurs after endovascular repair of thoracic aortic dissections because of continuing expansion of the true lumen. A suboptimal proximal landing zone, minimal oversizing, and lack of a healthy distal attachment site are unique factors affecting long-term stent graft stability. We used computational fluid dynamic techniques to analyze the biomechanical factors associated with stent graft remodeling in these patients. PATIENTS AND METHODS: A series of computational fluid dynamic models were constructed to investigate the biomechanical factors affecting the drag force on a thoracic stent graft. The resultant drag force as a net change of fluid momentum was calculated on the basis of varying three-dimensional geometry and deployment positions. A series of 12 patients with type B aortic dissections treated by thoracic stent graft and followed up for more than 12 months were then studied. Computed tomography transaxial images of each patient shortly after stent graft deployment and on subsequent follow-up were used to generate three-dimensional geometric models that were then fitted with a surface mesh. Computational fluid dynamic simulations were then performed on each stent graft model according to its geometric parameters to determine the actual change in drag force experienced by the stent graft as it remodels over time.
RESULTS: The drag force on the stent graft model increases linearly with its internal diameter and becomes highest when the deployment position is closer to the proximal arch. Aortic curvature is not a significant factor. Serial computed tomography scans of patients showed an increase in mean inlet area from 1030 mm(2) to 1140 mm(2), and mean outlet area from 586 mm(2) to 884 mm(2) (increase of 11% and 58%, respectively; P = .05, .01). These increases are associated with a change in resultant drag force on the stent graft from 21.0 N to 24.8 N (mean increase, 19.5%; range, 0%-63.2%; P = .002). There is a positive relationship between increase in drag force and increase in stent-graft area.
CONCLUSION: The drag force on thoracic stent grafts is high. A significant change in stent-graft diameter occurs after endovascular repair for type B dissections, which is associated with an increase in hemodynamic drag force. These stent grafts may be subjected to a higher risk of distal migration, and continuing surveillance is mandatory.

Entities:  

Mesh:

Year:  2008        PMID: 18644477     DOI: 10.1016/j.jvs.2008.03.050

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  7 in total

1.  Finite element analysis of helical flows in human aortic arch: a novel index.

Authors:  Cheng-Hung Lee; Kuo-Sheng Liu; Guan-Heng Jhong; Shih-Jung Liu; Ming-Yi Hsu; Chao-Jan Wang; Kuo-Chun Hung
Journal:  Biomicrofluidics       Date:  2014-04-09       Impact factor: 2.800

2.  The Necessity to Seal the Re-Entry Tears of Aortic Dissection After TEVAR: A Hemodynamic Indicator.

Authors:  Zhenfeng Li; Huanming Xu; Chlöe Harriet Armour; Yuze Guo; Jiang Xiong; Xiaoyun Xu; Duanduan Chen
Journal:  Front Bioeng Biotechnol       Date:  2022-03-31

3.  Computational Fluid Dynamics Study of Bifurcation Aneurysms Treated with Pipeline Embolization Device: Side Branch Diameter Study.

Authors:  Abraham Yik-Sau Tang; Wai-Choi Chung; Eric Tian-Yang Liu; Jie-Qiong Qu; Anderson Chun-On Tsang; Gilberto Ka-Kit Leung; Kar-Ming Leung; Alfred Cheuk-Hang Yu; Kwok-Wing Chow
Journal:  J Med Biol Eng       Date:  2015-06-30       Impact factor: 1.553

4.  A patient-specific study of type-B aortic dissection: evaluation of true-false lumen blood exchange.

Authors:  Duanduan Chen; Matthias Müller-Eschner; Hendrik von Tengg-Kobligk; David Barber; Dittmar Böckler; Rod Hose; Yiannis Ventikos
Journal:  Biomed Eng Online       Date:  2013-07-06       Impact factor: 2.819

Review 5.  A perspective review on numerical simulations of hemodynamics in aortic dissection.

Authors:  Wan Naimah Wan Ab Naim; Poo Balan Ganesan; Zhonghua Sun; Kok Han Chee; Shahrul Amry Hashim; Einly Lim
Journal:  ScientificWorldJournal       Date:  2014-02-03

6.  Predicting false lumen thrombosis in patient-specific models of aortic dissection.

Authors:  Claudia Menichini; Zhuo Cheng; Richard G J Gibbs; Xiao Yun Xu
Journal:  J R Soc Interface       Date:  2016-11       Impact factor: 4.118

7.  Endovascular repair during complex thoracic aortic dissection using a micropore stent graft: Midterm follow-up clinical outcomes.

Authors:  Zhe Wang; Cheng Wang; Fenghe Li; Yu Zhao
Journal:  Catheter Cardiovasc Interv       Date:  2019-08-13       Impact factor: 2.692

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.