Measurement of maximal static respiratory pressures at the mouth with different air leaks.

M Mayos1, J Giner, P Casan, J Sanchis.   

Abstract

The effect of two different circuit leaks on the measurement of maximal static inspiratory and expiratory pressures at the mouth (Pimax, Pemax) was assessed in 70 patients with respiratory disease. Patients were divided into three groups with similar anthropometric and spirometric characteristics. The first group (30 patients) had their Pmax measured with a leak of 2.0 mm internal diameter (ID) and 37 mm length (as proposed by T. Ringqvist) and repeated with a second leak of 1.0 mm ID and 15 mm length (as recommended by J. L. Clausen). The two measurements were done in random order. Measurements for the other two groups (20 patients each) were taken with one or another, the two leaks randomly alternated with no leak. Pimax measurements obtained with Ringqvist's leak were 17 percent (p less than 0.005) lower than those with Clausen's leak and 22 percent (p less than 0.005) lower than those with no leak. Pemax measurements performed with Ringqvist's leak in place were 11 percent (p less than 0.005) lower than those with Clausen's leak and 11 percent (p less than 0.005) lower than those obtained with no leak. The comparison between Clausen's leak and no leak showed no statistically significant difference. We conclude that whenever the effect of pressure generated in the mouth is to be avoided in the measurement of respiratory Pmax, a leak of the size proposed by Ringqvist is to be preferred.

Entities:  

Mesh:

Year:  1991        PMID: 1864106     DOI: 10.1378/chest.100.2.364

Source DB:  PubMed          Journal:  Chest        ISSN: 0012-3692            Impact factor:   9.410


  2 in total

Review 1.  Reference values for maximal inspiratory pressure: a systematic review.

Authors:  Isabela M B Sclauser Pessoa; Verônica Franco Parreira; Guilherme A F Fregonezi; A William Sheel; Frank Chung; W Darlene Reid
Journal:  Can Respir J       Date:  2013-10-17       Impact factor: 2.409

2.  Effects of transcutaneous electrical nerve stimulation on pain, walking function, respiratory muscle strength and vital capacity in kidney donors: a protocol of a randomized controlled trial.

Authors:  Thiago Tafarel Galli; Luciana Dias Chiavegato; Nathália Risso Santiago; Richard Eloin Liebano
Journal:  BMC Nephrol       Date:  2013-01-11       Impact factor: 2.388

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.