| Literature DB >> 18631375 |
Bernard Lambermont1, Alexandre Ghuysen, Nathalie Janssen, Philippe Morimont, Gary Hartstein, Paul Gerard, Vincent D'Orio.
Abstract
INTRODUCTION: Functional residual capacity (FRC) measurement is now available on new ventilators as an automated procedure. We compared FRC, static thoracopulmonary compliance (Crs) and PaO2 evolution in an experimental model of acute respiratory distress syndrome (ARDS) during a reversed, sequential ramp procedure of positive end-expiratory pressure (PEEP) changes to investigate the potential interest of combined FRC and Crs measurement in such a pathologic state.Entities:
Mesh:
Year: 2008 PMID: 18631375 PMCID: PMC2575573 DOI: 10.1186/cc6961
Source DB: PubMed Journal: Crit Care ISSN: 1364-8535 Impact factor: 9.097
Figure 1Time course of functional residual capacity (FRC) and static compliance of the respiratory system (Crs) during a decremental positive end-expiratory pressure (PEEP) trial after acute respiratory distress syndrome (ARDS) induction by oleic acid injection. Measurements were obtained 120 min after the oleic acid injection (ARDS) at a PEEP of 5 cm H2O and during a decremental PEEP trial from 20 to 0 cm H2O by steps of 5 cm H2O (ARDS20, ARDS15, ARDS10, ARDS5, ARDS0). Each PEEP level was maintained for 15 min before a set of measurements to allow for haemodynamic and respiratory stabilisation. § p < 0.05 vs ARDS (Crs); * p < 0.05 vs ARDS (FRC).
Time course of haemodynamic parameters, inspiratory plateau airway pressure, arterial blood pH, PaO2 and PaCO2 after oleic acid injection and during a decremental PEEP trial from 20 to 0 cm H2O.
| ARDS | 129 ± 3 | 75 ± 16 | 7.32 ± 0.06 | 52 ± 5 | 54 ± 5 | 20 ± 1 |
| ARDS20 | 132 ± 4 | 66 ± 13 | 7.33 ± 0.04 | 172 ± 15* | 54 ± 3 | 36 ± 3* |
| ARDS15 | 134 ± 4 | 73 ± 13 | 7.38 ± 0.02 | 175 ± 17* | 54 ± 2 | 26 ± 1* |
| ARDS10 | 145 ± 3* | 75 ± 9 | 7.28 ± 0.05 | 89 ± 19* | 56 ± 5 | 23 ± 2 |
| ARDS5 | 165 ± 10* | 118 ± 9* | 7.33 ± 0.02 | 52 ± 4* | 59 ± 2 | 22 ± 2 |
| ARDS0 | 153 ± 7* | 101 ± 5* | 7.3 ± 0.03 | 52 ± 4* | 62 ± 3* | 21 ± 2 |
Data are presented as mean ± standard error of the mean. Measurements were obtained 120 min after oleic acid injection at a PEEP of 5 cm H2O (ARDS) and during a decremential PEEP trial (ARDS20, ARDS15, ARDS10, ARDS5, ARDS0). Each PEEP level was maintained for 15 min before each measurement to allow for haemodynamic and respiratory stabilisation. ARDS, acute respiratory distress syndrome; HR, heart rate; mean AP, mean systemic arterial pressure; PEEP, positive end-expiratory pressure. * p < 0.05 versus ARDS
Figure 2Correlation between PaO2 and functional residual capacity (FRC).
Figure 3Correlation between PaO2 and static compliance of the respiratory system (Crs).
Figure 4Correlation between functional residual capacity (FRC) and static compliance of the respiratory system (Crs).